MB Sample ID: SA021331

Local Sample ID:26mar12_43-r002.d
Subject ID:SU000442
Subject Type:Animal
Subject Species:Homo sapiens
Taxonomy ID:9606
Species Group:Human

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN000663 AN000664 AN000665 AN000666
Analysis type MS MS MS MS
Chromatography type HILIC HILIC Reversed phase Reversed phase
Chromatography system Agilent 6220 Agilent 6220 Agilent 6220 Agilent 6220
Column None None None None
MS Type ESI ESI ESI ESI
MS instrument type TOF TOF TOF TOF
MS instrument name Agilent 6220 TOF Agilent 6220 TOF Agilent 6220 TOF Agilent 6220 TOF
Ion Mode POSITIVE NEGATIVE POSITIVE NEGATIVE
Units intensity intensity intensity intensity

Chromatography:

Chromatography ID:CH000480
Chromatography Summary:The liquid chromatography platform consisted of an Acquity UPLC system (Waters, Milford, MA). Plasma metabolite separation was achieved using both hydrophilic interaction chromatography (ethylene-bridged hybrid 2.1 × 150 mm, 1.7 μm; Waters) and reversed-phase liquid chromatography C18 (high-strength silica 2.1 × 150 mm, 1.8 μm; Waters). For each column, the run time was 20 min at a flow rate of 400 μL/min. Reverse-phase chromatography was performed using 99% solvent A (5 mmol/L NH4 acetate, 0.1% formic acid, and 1% acetonitrile) to 100% solvent B (95% acetonitrile with 0.1% formic acid). The gradient was 0 min, 0% B; 1 min, 0% B; 3 min, 5% B; 13.0 min, 100% B; 16 min, 100% B; 16.5 min, 0% B; and 20 min, 0% B. Other LC parameters were injection volume 5 μL and column temperature 50°C. Each sample was injected in triplicate with blank injections between each sample. Quality controls and standards were run at the beginning and end of the sequence.
Instrument Name:Agilent 6220
Column Name:None
Column Temperature:50
Flow Gradient:0 min, 0% B; 1 min, 0% B; 3 min, 5% B; 13.0 min, 100% B; 16 min, 100% B; 16.5 min, 0% B; and 20 min, 0% B.
Flow Rate:400 µL/min
Solvent A:1% acetonitrile/99% water; 0.1% formic acid; 5 mM ammonium acetate
Solvent B:95% acetonitrile/5% water; 0.1% formic acid
Chromatography Type:HILIC
  
Chromatography ID:CH000481
Chromatography Summary:The liquid chromatography platform consisted of an Acquity UPLC system (Waters, Milford, MA). Plasma metabolite separation was achieved using both hydrophilic interaction chromatography (ethylene-bridged hybrid 2.1 × 150 mm, 1.7 μm; Waters) and reversed-phase liquid chromatography C18 (high-strength silica 2.1 × 150 mm, 1.8 μm; Waters). For each column, the run time was 20 min at a flow rate of 400 μL/min. Reverse-phase chromatography was performed using 99% solvent A (5 mmol/L NH4 acetate, 0.1% formic acid, and 1% acetonitrile) to 100% solvent B (95% acetonitrile with 0.1% formic acid). The gradient was 0 min, 0% B; 1 min, 0% B; 3 min, 5% B; 13.0 min, 100% B; 16 min, 100% B; 16.5 min, 0% B; and 20 min, 0% B. Other LC parameters were injection volume 5 μL and column temperature 50°C. Each sample was injected in triplicate with blank injections between each sample. Quality controls and standards were run at the beginning and end of the sequence.
Instrument Name:Agilent 6220
Column Name:None
Column Temperature:50
Flow Gradient:0 min, 0% B; 1 min, 0% B; 3 min, 5% B; 13.0 min, 100% B; 16 min, 100% B; 16.5 min, 0% B; and 20 min, 0% B.
Flow Rate:400 µL/min
Solvent A:99% water/1% acetonitrile; 0.1% formic acid; 5 mM ammonium acetate
Solvent B:95% acetonitrile/5% water; 0.1% formic acid
Chromatography Type:Reversed phase