MB Sample ID: SA125922

Local Sample ID:PlasmaIS0_54
Subject ID:SU001567
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002475
Analysis type MS
Chromatography type Reversed phase
Chromatography system Waters Acquity
Column Waters Acquity CSH C18 (100 x 2.1mm,1.7um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode POSITIVE
Units Peak area

Chromatography:

Chromatography ID:CH001813
Chromatography Summary:LC-MS lipid analysis was performed on an Ultimate 3000 High-Performance UPLC. Chromatography separation was achieved with an Acquity UPLC CSH column [1.7 μm, 100 × 2.1 mm, (Waters Corporation, Milford, MA)] under 55°C with a flow rate of 0.4 ml/min. Mobile phase A was composed of ultrapure water/acetonitrile 40:60 (v/v), 10 mM ammonium formate and 0.1% formic acid. Mobile phase B contained ACN/IPA 10:90 (v/v) with 10 mM ammonium formate and 0.1% formic acid. The LC gradient started with 40% mobile phase B and raised to 43% mobile phase B in 2 min. The percentage of mobile phase B raised to 50% in the next 0.1 min and increased to 54% in next 9.9 min. Mobile phase B raised to 70% in 0.1 min and increased to 99% in 5.9 min and maintained at 99% for 1 min. Then the percentage of mobile phase B went back to 40% in 0.1 min and the system was equilibrated for 3.9 min before the next run started.
Instrument Name:Waters Acquity
Column Name:Waters Acquity CSH C18 (100 x 2.1mm,1.7um)
Column Temperature:under 55°C
Flow Rate:0.4 ml/min
Solvent A:40% water/60% acetonitrile; 0.1% formic acid; 10 mM ammonium formate
Solvent B:10% acetonitrile/90% isopropanol; 0.1% formic acid; 10 mM ammonium formate
Chromatography Type:Reversed phase