Return to study ST002499 main page
MB Sample ID: SA250395
Local Sample ID: | 8-4 |
Subject ID: | SU002596 |
Subject Type: | Mammal |
Subject Species: | Mus musculus |
Taxonomy ID: | 10090 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004106 |
---|---|
Analysis type | MS |
Chromatography type | Reversed phase |
Chromatography system | Thermo Dionex Ultimate 3000 |
Column | Waters XBridge C18 (50 x 2.1mm,3um) |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo Exactive Plus Orbitrap |
Ion Mode | NEGATIVE |
Units | peak area normalized to IS and cell number |
Chromatography:
Chromatography ID: | CH003041 |
Chromatography Summary: | The sample run order was randomized to reduce bias from instrument drift. 10 µl sample was subjected to LC-MS analysis on a Exactive Plus Orbitrap mass spectrometer (Thermo Fisher Scientific) coupled to an Ultimate 3000 UHPLC system (Thermo Fisher Scientific). Reversed-phase chromatography mode was used to separate compounds on a Xselect C18 HSS column (Waters) with solvent A (97:3 water:methanol (v/v), 10 mM tributylamine, and 15 mM acetic acid ) and solvent B (methanol). The flow rate was 200 µl/min, and the total run time was 25 min. The gradient was 0 min, 0% B; 5 min, 20% B; 7.5 min, 55% B; 15 min, 65% B; 17.5 min, 95% B; and 21 min, 0% B. The mass spectrometer was operated in a negative-ion mode at a resolution of 140,000 at m/z 200 and with a scan range of 85 to 1000 m/z. |
Instrument Name: | Thermo Dionex Ultimate 3000 |
Column Name: | Waters XBridge C18 (50 x 2.1mm,3um) |
Column Temperature: | standard |
Flow Gradient: | 0 min, 0% B; 5 min, 20% B; 7.5 min, 55% B; 15 min, 65% B; 17.5 min, 95% B; and 21 min, 0% B. |
Flow Rate: | 200 µl/min |
Solvent A: | 97:3 water:methanol (v/v), 10 mM tributylamine, and 15 mM acetic acid |
Solvent B: | methanol |
Chromatography Type: | Reversed phase |