Return to study ST002763 main page

MB Sample ID: SA291575

Local Sample ID:A5_CE-MS
Subject ID:SU002870
Subject Type:Bacteria
Subject Species:Haemophilus influenzae
Genotype Strain:Rd KW20
Species Group:Bacteria

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004492 AN004493 AN004494 AN004495 AN004496 AN004497
Analysis type MS MS MS MS MS MS
Chromatography type GC CE Reversed phase Reversed phase Reversed phase Reversed phase
Chromatography system Agilent 7890B Agilent 7100 CE Agilent 1290 Infinity Agilent 1290 Infinity Agilent 1290 Infinity Agilent 1290 Infinity
Column Agilent DB5-MS (30m x 0.25mm, 0.25um) fused silica capillary (70 cm; 50 um i.d.), Agilent Technologies Agilent InfinityLab Poroshell 120 EC-C18 (100 x 3mm,2.7um) Agilent InfinityLab Poroshell 120 EC-C18 (100 x 3mm,2.7um) Agilent InfinityLab Poroshell 120 EC-C18 (100 x 3mm,2.7um) Agilent InfinityLab Poroshell 120 EC-C18 (100 x 3mm,2.7um)
MS Type EI ESI ESI ESI ESI ESI
MS instrument type QTOF TOF QTOF QTOF QTOF QTOF
MS instrument name Agilent 7250 Agilent 6224 TOF Agilent 6545 QTOF Agilent 6545 QTOF Agilent 6545 QTOF Agilent 6545 QTOF
Ion Mode UNSPECIFIED POSITIVE POSITIVE NEGATIVE POSITIVE NEGATIVE
Units raw abundance raw abundance raw abundance raw abundance raw abundance

Chromatography:

Chromatography ID:CH003377
Chromatography Summary:Chromatography was performed according to DOI: 10.1385/1-59745-003-0:439
Instrument Name:Agilent 7890B
Column Name:Agilent DB5-MS (30m x 0.25mm, 0.25um)
Column Temperature:According to temperature gradient. DOI: 10.1385/1-59745-003-0:439
Flow Gradient:N/A for GC
Flow Rate:0.85 mL/min
Internal Standard:Tricosane
Sample Injection:2 uL
Solvent A:N/A for GC
Solvent B:N/A for GC
Chromatography Type:GC
Solvent C:N/A for GC
  
Chromatography ID:CH003378
Chromatography Summary:Separation was performed according to DOI: 10.1128/msphere.00184-22
Instrument Name:Agilent 7100 CE
Column Name:fused silica capillary (70 cm; 50 um i.d.), Agilent Technologies
Column Temperature:RT
Flow Gradient:N/A for CE-MS
Flow Rate:N/A for CE-MS
Internal Standard:Methionine sulfone
Sample Injection:hydrodynamic injection: 50 mbar for 100 s
Solvent A:N/A for CE-MS
Solvent B:N/A for CE-MS
Capillary Voltage:30 kV
Running Buffer:1M HCOOH in MeOH:H2O 1:9 (v/v) (BGE)
Chromatography Type:CE
Solvent C:N/A for CE-MS
  
Chromatography ID:CH003379
Chromatography Summary:Chromatography was performed according to DOI: 10.1128/msphere.00184-22
Instrument Name:Agilent 1290 Infinity
Column Name:Agilent InfinityLab Poroshell 120 EC-C18 (100 x 3mm,2.7um)
Column Temperature:50 ÂșC
Flow Gradient:70%B held 1 min; then increase up to 86 %B at 3.5 min; then hold 10 min; Next, incerase in %B up to 100 %B at 11 min; then, held for 6 min. Then, switched to initial conditions (1.84 min of re-equilibration time)
Flow Rate:0.6 mL/min
Sample Injection:2 uL (ESI+, MS1, ITMS with default inj. volume) and 5 uL (ESI-, MS1, ITMS with default inj. volume), 7 uL (ESI+ ITMS with increased inj. volume), 10 uL (ESI- ITMS with increased inj. volume)
Solvent A:10 mM Ammonium acetate and 0.2 mM ammonium fluoride in 9:1 H2O:MeOH (v/v)
Solvent B:10 mM Ammonium acetate and 0.2 mM ammonium fluoride in 2:3:5 ACN:MeOH:IPA (v/v)
Chromatography Type:Reversed phase
Solvent C:N/A