Summary of Study ST001115

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000746. The data can be accessed directly via it's Project DOI: 10.21228/M87X0C This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001115
Study TitleGrowth cone-enriched lipidome of embryonic to early postnatal mouse brain
Study SummaryA growth cone (GC) is a part of a neuron considered a hub for axon growth, motility and guidance functions. Unraveling the molecular composition of GCs and events through which active GCs transition to terminal synapses is of importance to developmental and regenerative neuroscience research. Here, we present a dataset on the lipid profiling of the growth cone-enriched fractions derived from E18, P0, P3, P6 and P9 C57BL/6J mouse brains. For comparison, we analyzed non-growth cone membranes.
Institute
University of Miami
DepartmentOphthalmology, Bascom Palmer Eye Institute
LaboratorySanjoy K. Bhattacharya
Last NameBhattacharya
First NameSanjoy
Address1638 NW 10th Ave, Miami, FL 33136
Emailsbhattacharya@med.miami.edu
Phone+1 (305) 4824103
Submit Date2018-11-26
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2019-01-16
Release Version1
Sanjoy Bhattacharya Sanjoy Bhattacharya
https://dx.doi.org/10.21228/M87X0C
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN001811
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Accela 600
Column Thermo Acclaim C30 (150 x 2.1mm,3um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode UNSPECIFIED
Units main area

Chromatography:

Chromatography ID:CH001278
Chromatography Summary:Reversed phase chromatographic separation utilized Accela Autosampler (Thermo), Accela 600 pump (Thermo) and Acclaim C30 column: 3 µm, 2.1x150 mm (Thermo). The column temperature was maintained at 30˚C and tray temperature at 20˚C. Solvent A was composed of 10 mM ammonium acetate (LC-MS grade) in 60:40 methanol:water (LC-MS grade) with 0.2% formic acid (FA; LC-MS grade). Solvent B was composed of 10 mM ammonium acetate with 60:40 methanol:chloroform with 0.2% FA. The flow rate was 260 µL/min and injection volume was 10 µL. The gradient was 35-100% solvent B over 13.0 min, 100% solvent B over 13.0-13.8 min, 100-35% solvent B over 13.8-14.5 min, 35% solvent B over 14.5-18.0 min, 0% solvent B over 18.0-20.0 min.
Instrument Name:Thermo Accela 600
Column Name:Thermo Acclaim C30 (150 x 2.1mm,3um)
Column Temperature:30
Flow Gradient:35-100% solvent B over 13.0 min, 100% solvent B over 13.0-13.8 min, 100-35% solvent B over 13.8-14.5 min, 35% solvent B over 14.5-18.0 min, 0% solvent B over 18.0-20.0 min
Flow Rate:260 µL/min
Solvent A:60% methanol/40% water; 0.2% formic acid; 10 mM ammonium acetate
Solvent B:60% methanol/40% chloroform; 0.2% formic acid; 10 mM ammonium acetate
Chromatography Type:Reversed phase