This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000813. The data can be accessed directly via it's Project DOI: 10.21228/M8KX27 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001210
Study TitleComprehensive UHPLC-MS/MS lipidomics profiling to study effects of betulin on keratinocytes
Study SummaryLipidomics analysis of betulin in human primary keratinocytes to monitor alterations in the lipid profiles induced by treatment with betulin
Institute
Eberhard Karls University of Tübingen
DepartmentInstitute of Pharmaceutical Sciences
LaboratoryPharmaceutical (Bio-)Analysis
Last NameCalderon
First NameCarlos
AddressAuf der Morgenstelle 8 (Haus B), D-72076, Tuebingen, baden württemberg, 72076, Germany
Emailcarlos.calderon@uni-tuebingen.de
Phone+49 (0)7071 29 74009
Submit Date2019-07-05
Num Groups2
Total Subjects20
Study CommentsEberhard-Karls-University Tuebingen
Raw Data AvailableYes
Raw Data File Type(s)wiff
Analysis Type DetailLC-MS
Release Date2019-09-23
Release Version1
Carlos Calderon Carlos Calderon
https://dx.doi.org/10.21228/M8KX27
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002014 AN002015
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Agilent 1290 Agilent 1290
Column Waters ACQUITY UPLC CSH C18 (100 x 2.1mm,1.7um,130Å) Waters ACQUITY UPLC CSH C18 (100 x 2.1mm,1.7um,130Å)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name ABI Sciex 5600 TripleTOF ABI Sciex 5600 TripleTOF
Ion Mode POSITIVE NEGATIVE
Units Normalized intensities Normalized intensities

Chromatography:

Chromatography ID:CH001457
Chromatography Summary:Analyses were performed by Agilent 1290 Series UHPLC instrument coupled to Sciex TripleTOF 5600+ MS with duospray source and Pal HTC-XS autosampler from CTC. Positive and negative ESI ionization were used in separate LC-MS runs. Conditions: Acquity UPLC CSH C18 (130Å, 1.7 µm, 2.1 mm X 100 mm) column was utilized with Acquity UPLC CSH C18 VanGuard pre-column (130Å, 1.7 µm, 2.1 mm X 5 mm). The mobile phase was composed of 10 mM ammonium formate and 0.1 % formic acid dissolved in 60:40 ACN:H2O (v/v) (A) and 90:10 (v/v) IPA:ACN (B).
Methods Filename:ccalcas_LC-MS.pdf
Instrument Name:Agilent 1290
Column Name:Waters ACQUITY UPLC CSH C18 (100 x 2.1mm,1.7um,130Å)
Column Temperature:65
Flow Rate:600 uL/min
Solvent A:100% water; 5 mM ammonium acetate, pH 9.9
Solvent B:100% acetonitrile
Chromatography Type:Reversed phase