This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000830. The data can be accessed directly via it's Project DOI: 10.21228/M8D68Q This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001242
Study TitleGlobal Metabolic Analysis Trisomy 21 - Cohort 3, CSF
Study Summaryglobal metabolic analysis of CSF from individuals with and without trisomy 21.
Institute
University of Colorado Denver
Last NameCulp-Hill
First NameRachel
Address12801 E 17th Ave L18-9403D, Aurora, Colorado, 80045, USA
Emailrachel.hill@cuanschutz.edu
Phone303-724-5798
Submit Date2019-08-13
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2019-08-21
Release Version1
Rachel Culp-Hill Rachel Culp-Hill
https://dx.doi.org/10.21228/M8D68Q
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002063 AN002064
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Thermo Vanquish Thermo Vanquish
Column Phenomenex Kinetex C18 (150 x 2.1mm,2.6um) Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units relative abundance relative abundance

Chromatography:

Chromatography ID:CH001501
Chromatography Summary:UHPLC-MS metabolomics analyses were performed using a Vanquish UHPLC system coupled online to a Q Exactive mass spectrometer (Thermo Fisher, Bremen, Germany). 20uL of supernatant was injected for plasma and CSF samples, and 10uL of supernatant was injected for cell samples. Samples were resolved over a Kinetex C18 column (2.1x150 mm, 1.7μm; Phenomenex, Torrance, CA, USA) at 45°C using a 5-minute gradient at 450μL/minute from 5-95% B (A: water/0.1% formic acid; B: acetonitrile/0.1% formic acid) for positive mode. To monitor possible technical variability, aliquots of each of the individual samples were combined to make technical replicates, which were run after every 15 samples. Additionally, in each experiment, several lysis solution aliquots were run as blanks for artifact identification.
Methods Filename:5MMpos_method.pdf
Instrument Name:Thermo Vanquish
Column Name:Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
Column Temperature:45
Flow Gradient:5-95% B
Flow Rate:450uL/min
Solvent A:100% water; 0.1% formic acid
Solvent B:100% acetonitrile; 0.1% formic acid
Chromatography Type:Reversed phase
  
Chromatography ID:CH001502
Chromatography Summary:UHPLC-MS metabolomics analyses were performed using a Vanquish UHPLC system coupled online to a Q Exactive mass spectrometer (Thermo Fisher, Bremen, Germany). 20uL of supernatant was injected for each sample. Samples were resolved over a Kinetex C18 column (2.1x150 mm, 1.7μm; Phenomenex, Torrance, CA, USA) at 45°C using a 5-minute gradient at 450μL/minute from 0-100% B (A: 95% water/5% acetonitrile, 1mM NH4OAc; B: 95% acetonitrile/5% water, 1mM NH4OAc) for negative mode. To monitor possible technical variability, aliquots of each of the individual samples were combined to make technical replicates, which were run after every 15 samples. Additionally, in each experiment, several lysis solution aliquots were run as blanks for artifact identification.
Methods Filename:5MMneg_method.pdf
Instrument Name:Thermo Vanquish
Column Name:Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
Column Temperature:45
Flow Gradient:0-100% B
Flow Rate:450uL/min
Solvent A:95% water/5% acetonitrile; 1 mM ammonium acetate
Solvent B:95% acetonitrile/5% water; 1 mM ammonium acetate
Chromatography Type:Reversed phase