Summary of Study ST001493

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001011. The data can be accessed directly via it's Project DOI: 10.21228/M8111X This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001493
Study TitleDynamic binning peak detection and assessment of various lipidomics liquid chromatography-mass spectrometry pre-processing platforms
Study SummaryLiquid chromatography-mass spectrometry (LC-MS) based lipidomics generate a large dataset, which requires high-performance data pre-processing tools for their interpretation such as XCMS, mzMine and Progenesis. These pre-processing tools rely heavily on accurate peak detection, which depends on setting the peak detection mass tolerance (PDMT) properly. The PDMT is usually set with a fixed value in either ppm or Da units. However, this fixed value may result in duplicates or missed peak detection. Therefore, we developed the dynamic binning method for accurate peak detection, which takes into account the peak broadening described by well-known physics laws of ion separation and set dynamically the value of PDMT as a function of m/z. Namely, in our method, the PDMT is proportional to for FTICR, to for Orbitrap, to m/z for Q-TOF and is a constant for Quadrupole mass analyzer, respectively. The dynamic binning method was implemented in XCMS. Our further goal was to compare the performance of different lipidomics pre-processing tools to find differential compounds. We have generated set samples with 43 lipids internal standards differentially spiked to aliquots of one human plasma lipid sample using Orbitrap LC-MS/MS. The performance of the various pipelines using aligned parameter sets was quantified by a quality score system which reflects the ability of a pre-processing pipeline to detect differential peaks spiked at various concentration levels. The quality score indicates that the dynamic binning method improves the performance of XCMS (maximum p-value 9.8·10-3 of two-sample Wilcoxon test). The modified XCMS software was further compared with mzMine and Progenesis. The results showed that modified XCMS and Progenesis had a similarly good performance in the aspect of finding differential compounds. In addition, Progenesis shows lower variability as indicated by lower CVs, followed by XCMS and mzMine. The lower variability of Progenesis improve the quantification, however, provide an incorrect quantification abundance order of spiked-in internal standards.
Institute
University of Groningen
Last NamePéter
First NameHorvatovich
AddressAntonius Deusinglaan 1, 9713 AV Groningen, The Netherlands
Emailp.l.horvatovich@rug.nl
Phone+31 (0)50 363 3341
Submit Date2020-09-25
Num Groups6
Total Subjects1
Study CommentsDifferent concentrations of lipid standard mixture were added to the plasma lipid extract aliquots
PublicationsUnder review
Raw Data AvailableYes
Raw Data File Type(s)mzML
Analysis Type DetailLC-MS
Release Date2020-10-13
Release Version1
Horvatovich Péter Horvatovich Péter
https://dx.doi.org/10.21228/M8111X
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002475
Analysis type MS
Chromatography type Reversed phase
Chromatography system Waters Acquity
Column Waters Acquity CSH C18 (100 x 2.1mm,1.7um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode POSITIVE
Units Peak area

Chromatography:

Chromatography ID:CH001813
Chromatography Summary:LC-MS lipid analysis was performed on an Ultimate 3000 High-Performance UPLC. Chromatography separation was achieved with an Acquity UPLC CSH column [1.7 μm, 100 × 2.1 mm, (Waters Corporation, Milford, MA)] under 55°C with a flow rate of 0.4 ml/min. Mobile phase A was composed of ultrapure water/acetonitrile 40:60 (v/v), 10 mM ammonium formate and 0.1% formic acid. Mobile phase B contained ACN/IPA 10:90 (v/v) with 10 mM ammonium formate and 0.1% formic acid. The LC gradient started with 40% mobile phase B and raised to 43% mobile phase B in 2 min. The percentage of mobile phase B raised to 50% in the next 0.1 min and increased to 54% in next 9.9 min. Mobile phase B raised to 70% in 0.1 min and increased to 99% in 5.9 min and maintained at 99% for 1 min. Then the percentage of mobile phase B went back to 40% in 0.1 min and the system was equilibrated for 3.9 min before the next run started.
Instrument Name:Waters Acquity
Column Name:Waters Acquity CSH C18 (100 x 2.1mm,1.7um)
Column Temperature:under 55°C
Flow Rate:0.4 ml/min
Solvent A:40% water/60% acetonitrile; 0.1% formic acid; 10 mM ammonium formate
Solvent B:10% acetonitrile/90% isopropanol; 0.1% formic acid; 10 mM ammonium formate
Chromatography Type:Reversed phase