Summary of Study ST001760

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001127. The data can be accessed directly via it's Project DOI: 10.21228/M81D6M This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001760
Study TitleApplication of the redox metabolite detection method for mouse kidney
Study SummaryThis study was aimed at optimizing redox metabolites detection from mammalian tissues. Three different chromatographic conditions were compared as well as three different extraction conditions. This study was run on ZIC-pHILIC chromatography. This study is an independent replicate
Institute
Boston Children's Hospital, Harvard Medical School
DepartmentPathology
LaboratoryNaama Kanarek
Last NamePetrova
First NameBoryana
Address300 Longwood Av, Boston, MA, 2115, USA
Emailboryana.petrova@childrens.harvard.edu
Phone6173557433
Submit Date2021-04-21
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2021-05-17
Release Version1
Boryana Petrova Boryana Petrova
https://dx.doi.org/10.21228/M81D6M
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002867
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Vanquish
Column EMD Millipore ZIC HILIC (150 x 2.1mm,5um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode UNSPECIFIED
Units ppm

Chromatography:

Chromatography ID:CH002122
Chromatography Summary:ZIC-pHILIC chromatography: One ml of reconstituted sample was injected into a ZIC-pHILIC 150 × 2.1 mm (5 µm particle size) column (EMD Millipore) operated on a Vanquish™ Flex UHPLC Systems (Thermo Fisher Scientific, San Jose, CA). Chromatographic separation was achieved using the following conditions: buffer A was acetonitrile; buffer B was 20 mM ammonium carbonate, 0.1% ammonium hydroxide. Gradient conditions were: linear gradient from 20% to 80% B; 20–20.5 min: from 80% to 20% B; 20.5–28 min: hold at 20% B. The column oven and autosampler tray were held at 25 °C and 4 °C, respectively.
Instrument Name:Thermo Vanquish
Column Name:EMD Millipore ZIC HILIC (150 x 2.1mm,5um)
Column Temperature:25
Flow Gradient:linear gradient from 20% to 80% B; 20-20.5 min: from 80% to 20% B; 20.5-28 min: hold at 20% B.
Solvent A:100% acetonitrile
Solvent B:100% water; 20 mM ammonium carbonate; 0.1% ammonium hydroxide
Chromatography Type:HILIC
  logo