Summary of Study ST002223

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001418. The data can be accessed directly via it's Project DOI: 10.21228/M8F409 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST002223
Study TitleMetabolic profiling of mouse tissues and tissue interstitial fluids
Study SummaryTissue and tissue interstitial fluids was collected from mice of a hybrid C57BL/6J;129/SvJ background of about 12 weeks of age (8 in total) and used to profile the metabolic content. This is Part 6 of a study and the experimental number is MS56.
Institute
CECAD Research Center
Last NameYang
First NameMing
AddressJoseph-Stelzmann-Straße 26, Köln, Koeln, 50931, Germany
Emailming.yang@uni-koeln.de
Phone4922147884306
Submit Date2022-07-15
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2022-08-03
Release Version1
Ming Yang Ming Yang
https://dx.doi.org/10.21228/M8F409
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003632
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Dionex Ultimate 3000
Column SeQuant ZIC-pHILIC
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode UNSPECIFIED
Units peak area

Chromatography:

Chromatography ID:CH002687
Chromatography Summary:Chromatographic separation of metabolites was achieved using a Millipore Sequant ZIC-pHILIC analytical column (5 µm, 2.1 × 150 mm) equipped with a 2.1 × 20 mm guard column (both 5 mm particle size) with a binary solvent system. Solvent A was 20 mM ammonium carbonate, 0.05% ammonium hydroxide; Solvent B was acetonitrile. The column oven and autosampler tray were held at 40 °C and 4 °C, respectively. The chromatographic gradient was run at a flow rate of 0.200 mL/min as follows: 0–2 min: 80% B; 2-17 min: linear gradient from 80% B to 20% B; 17-17.1 min: linear gradient from 20% B to 80% B; 17.1-23 min: hold at 80% B. Samples were randomized and the injection volume was 5 µl. A pooled quality control (QC) sample was generated from an equal mixture of all individual samples and analysed interspersed at regular intervals.
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:SeQuant ZIC-pHILIC
Column Temperature:40
Flow Gradient:0-2 min: 80% B; 2-17 min: linear gradient from 80% B to 20% B; 17-17.1 min: linear gradient from 20% B to 80% B; 17.1-23 min: hold at 80% B
Flow Rate:0.200 mL/min
Solvent A:100% water; 20 mM ammonium carbonate; 0.05% ammonium hydroxide
Solvent B:100% acetonitrile
Chromatography Type:HILIC