Summary of Study ST002956

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001839. The data can be accessed directly via it's Project DOI: 10.21228/M80T6F This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST002956
Study TitleAutomated preparation of plasma lipids, metabolites, and proteins for LC/MS-based analysis of a high-fat diet in mice
Study TypeMethod Development
Study SummaryWe designed an automated liquid-liquid extraction method with minimal contamination and human intervention. This approach enables accurate and precise collection of metabolite, lipid fractions, and protein pellet from a small volume of mice plasma for multiomic analysis.
Institute
Calico Life Sciences
DepartmentDepartment of Mass Spectrometry-Technology Lab
LaboratoryMetabolomics Lab
Last NameVu
First NameNgoc
Address1130 Veterans BLVD
Emailngoc@calicolabs.com
Phone6504205430
Submit Date2023-07-06
Num Groups3
Total Subjects90
Num Males90
Raw Data AvailableYes
Raw Data File Type(s)mzML
Analysis Type DetailLC-MS
Release Date2024-07-29
Release Version1
Ngoc Vu Ngoc Vu
https://dx.doi.org/10.21228/M80T6F
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004853 AN004854 AN004855 AN004856
Analysis type MS MS MS MS
Chromatography type Reversed phase Reversed phase HILIC Ion pair
Chromatography system Thermo Vanquish Thermo Vanquish Thermo Vanquish Thermo Vanquish
Column Thermo Accucore C30 column (250 x 2.1 mm, 2.6 µm) Thermo Accucore C30 column (250 x 2.1 mm, 2.6 µm) SeQuant ZIC- pHILIC (150 x 2.1mm,5um) Agilent Zorbax Extend C18 column (150 x 2.1 mm, 1.8 µm)
MS Type ESI ESI ESI ESI
MS instrument type Orbitrap Orbitrap Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Plus Orbitrap Thermo Q Exactive Plus Orbitrap Thermo Q Exactive Plus Orbitrap Thermo Q Exactive Plus Orbitrap
Ion Mode POSITIVE NEGATIVE POSITIVE NEGATIVE
Units Log2(Top Peak Area) Log2(Top Peak Area) Log2(Top Peak Area) Log2(Top Peak Area)

Chromatography:

Chromatography ID:CH003664
Chromatography Summary:RPLC method for lipidomics
Instrument Name:Thermo Vanquish
Column Name:Thermo Accucore C30 column (250 x 2.1 mm, 2.6 µm)
Column Temperature:35
Flow Gradient:The column was equilibrated in 30% B for 7 minutes before injection. The gradient then increased to 43% B for another 7 minutes, followed by an increase to 65% B over 5 min and then slowly increasing to 70% B over 18 min. B was then increased to 88% for 1 min, followed by another slow increase to 95% over 20 min. Over the next 2 min, B was increased to 100%, followed by isocratic flow for another 2 min. After this step, the gradient was dropped to 30% B within 0.1 min and kept on isocratic flow for the next 5 min, resulting in a total run time of 67 min.
Flow Rate:200 uL/min
Solvent A:20 mM ammonium formate in [60:40] ACN:H2O, 0.25 µM medronic acid
Solvent B:20 mM ammonium formate in 90:10 IPA:MeCN, 0.25 µM medronic acid
Chromatography Type:Reversed phase
  
Chromatography ID:CH003665
Chromatography Summary:HILIC method for Metabolomics
Instrument Name:Thermo Vanquish
Column Name:SeQuant ZIC- pHILIC (150 x 2.1mm,5um)
Column Temperature:35
Flow Gradient:t = −6, 80% B; t = 0, 80% B; t = 2.5, 73% B; t = 5, 65% B, t = 7.5, 57% B; t = 10, 50% B; t = 15, 35% B; t = 20; 20% B; t = 22, 15% B; t = 22.5, 80% B; t = 24; 80% B.
Flow Rate:150uL/min
Solvent A:20 mM ammonium carbonate in water (pH 9.2)
Solvent B:100% MeCN
Chromatography Type:HILIC
  
Chromatography ID:CH003666
Chromatography Summary:RPLC method for Metabolomics
Instrument Name:Thermo Vanquish
Column Name:Agilent Zorbax Extend C18 column (150 x 2.1 mm, 1.8 µm)
Column Temperature:30
Flow Gradient:The gradient eluted isocratically in 0% B for 2.5 min, increased to 20% B over 2.5 min, held at 20% B for 2.5 min, increased to 55% B over 5.5 min, increased to 95% B over 2.5 minutes, maintained at 95% B for 3 min, then decreased to 0% B over 0.5 min, where it was held for 3 min, resulting in a total run time of 26 min.
Flow Rate:200 uL/min
Solvent A:10 mM tributylamine and 15 mM acetic acid in [97:3] H2O:MeOH at pH 4.95 with 0.25 µM medronic acid
Solvent B:100% MeOH
Chromatography Type:Ion pair