Summary of Study ST003604
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR002231. The data can be accessed directly via it's Project DOI: 10.21228/M87R79 This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST003604 |
Study Title | Investigation of global metabolites and E coli-derived metabolites following engulfment of dead bacteria by wild type or RagAGTP-expressing bone marrow derived macrophages. |
Study Summary | Global metabolomics and 13C-tracing analysis of wild type or RagAGTP-expressing bone marrow derived macrophages exposed to heat-killed uniformly 13C-labelled E. coli in a trans-well system for 6h or 18h. |
Institute | University of Colorado Anschutz Medical Campus |
Last Name | Haines |
First Name | Julie |
Address | 12801 E 17th Ave, Room 1303, Aurora, Colorado, 80045, USA |
julie.haines@cuanschutz.edu | |
Phone | 3037243339 |
Submit Date | 2024-11-25 |
Raw Data Available | Yes |
Raw Data File Type(s) | raw(Thermo) |
Analysis Type Detail | LC-MS |
Release Date | 2024-12-27 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN005921 | AN005922 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | Reversed phase | Reversed phase |
Chromatography system | Thermo Vanquish | Thermo Vanquish |
Column | Phenomenex Kinetex C18 (150 x 2.1mm,1.7um) | Phenomenex Kinetex C18 (150 x 2.1mm,1.7um) |
MS Type | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap | Thermo Q Exactive Orbitrap |
Ion Mode | NEGATIVE | POSITIVE |
Units | peak area | peak area |
Chromatography:
Chromatography ID: | CH004496 |
Chromatography Summary: | Negative C18 |
Instrument Name: | Thermo Vanquish |
Column Name: | Phenomenex Kinetex C18 (150 x 2.1mm,1.7um) |
Column Temperature: | 45 |
Flow Gradient: | 0-0.5 min 0% B, 0.5-1.1 min 0-100% B, 1.1-2.75 min hold at 100% B, 2.75-3 min 100-0% B, 3-5 min hold at 0% B |
Flow Rate: | 450 uL/min |
Sample Injection: | 6 uL |
Solvent A: | 95% water/5% acetonitrile; 1 mM ammonium acetate |
Solvent B: | 95% acetonitrile/5% water; 1 mM ammonium acetate |
Chromatography Type: | Reversed phase |
Chromatography ID: | CH004497 |
Chromatography Summary: | Positive C18 |
Instrument Name: | Thermo Vanquish |
Column Name: | Phenomenex Kinetex C18 (150 x 2.1mm,1.7um) |
Column Temperature: | 45 |
Flow Gradient: | 0-0.5 min 5% B, 0.5-1.1 min 5-95% B, 1.1-2.75 min hold at 95% B, 2.75-3 min 95-5% B, 3-5 min hold at 5% B |
Flow Rate: | 450 uL/min |
Sample Injection: | 6 uL |
Solvent A: | 100% water; 0.1% formic acid |
Solvent B: | 100% acetonitrile; 0.1% formic acid |
Chromatography Type: | Reversed phase |