Summary of Study ST000069

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000066. The data can be accessed directly via it's Project DOI: 10.21228/M8QG6H This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Show all samples  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files
Study IDST000069
Study TitleMetabolomic profiling of influenza: a 2009 pandemic H1N1 influenza in lean and obese mice (via blood and tissue)
Study SummaryLung samples were obtained at 15 weeks from 56 mice that had received either chow, a low fat diet, or a high fat diet and that were uninfected, 4 days post-infection, or 8 days post-infection.
Institute
University of North Carolina
DepartmentSystems and Translational Sciences
LaboratorySumner Lab
Last NameSumner
First NameSusan
AddressEastern Regional Comprehensive Metabolomics Resource Core, UNC Nutrition Research Institute, 500 Laureate Way, Kannapolis, NC, 28081
Emailsusan_sumner @unc.edu
Phone704-250-5066
Submit Date2014-06-14
Num Groups9
Total Subjects55
Raw Data AvailableNo
Raw Data File Type(s)raw(Waters)
Analysis Type DetailLC-MS
Release Date2015-06-01
Release Version1
Susan Sumner Susan Sumner
https://dx.doi.org/10.21228/M8QG6H
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN000107 AN000108
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Waters Acquity Waters Acquity
Column Waters Acquity HSS T3 (100 x 2.1mm,1.8um) Waters Acquity HSS T3 (100 x 2.1mm,1.8um)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name Waters Synapt G2 QTOF Waters Synapt G2 QTOF
Ion Mode POSITIVE NEGATIVE
Units Normalized abundance Normalized abundance

MS:

MS ID:MS000083
Analysis ID:AN000107
Instrument Name:Waters Synapt G2 QTOF
Instrument Type:QTOF
MS Type:ESI
Ion Mode:POSITIVE
Capillary Temperature:110 C
Capillary Voltage:3.2 kV
Collision Energy:4
Helium Flow:180
Ionization:ES+
Mass Accuracy:10 ppm
Source Temperature:110 C
Dataformat:Continuum
Desolvation Gas Flow:400 L/Hr
Desolvation Temperature:400 C
Resolution Setting:18000
Scan Range Moverz:50-1000 m/s
Scanning Cycle:1 s
Tube Lens Voltage:75
  
MS ID:MS000084
Analysis ID:AN000108
Instrument Name:Waters Synapt G2 QTOF
Instrument Type:QTOF
MS Type:ESI
Ion Mode:NEGATIVE
Capillary Temperature:110 C
Capillary Voltage:2.8 kV
Collision Energy:4
Helium Flow:180
Ionization:ES-
Source Temperature:110 C
Dataformat:Continuum
Desolvation Gas Flow:400 L/Hr
Desolvation Temperature:400 C
Resolution Setting:18000
Scan Range Moverz:50-1000 m/z
Scanning Cycle:1 s
Tube Lens Voltage:74
  logo