Summary of Study ST000221

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000181. The data can be accessed directly via it's Project DOI: 10.21228/M8R30R This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST000221
Study TitleNormal plasma cells,Low proliferation multiple myeloma and High proliferation multiple myeloma cells
Study Typedifferential metabolomics
Study SummaryCD138 sorted bone marrow plasma cell were obtained from a normal patient, a multiple myelow with slow proliferation and a multiple myelow with rapid proliferation.
Institute
Mayo Clinic
DepartmentHematology
Last NameGonsalves
First NameWilliam
EmailDasari.Surendra@mayo.edu
Submit Date2015-06-20
Num Groups1
Total Subjects1
Raw Data AvailableYes
Raw Data File Type(s)d
Uploaded File Size73 G
Analysis Type DetailLC-MS
Release Date2016-06-18
Release Version1
William Gonsalves William Gonsalves
https://dx.doi.org/10.21228/M8R30R
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN000327 AN000328 AN000329 AN000330
Analysis type MS MS MS MS
Chromatography type Reversed phase Reversed phase HILIC HILIC
Chromatography system Waters Acquity Waters Acquity Waters Acquity Waters Acquity
Column Waters high-strength silica (150 x 2.1mm,1.8um) Waters high-strength silica (150 x 2.1mm,1.8um) Waters ethylene-bridged hybrid (150 x 2.1mm,1.7um) Waters ethylene-bridged hybrid (150 x 2.1mm,1.7um)
MS Type ESI ESI ESI ESI
MS instrument type TOF TOF TOF TOF
MS instrument name Agilent 6220 TOF Agilent 6220 TOF Agilent 6220 TOF Agilent 6220 TOF
Ion Mode POSITIVE NEGATIVE POSITIVE NEGATIVE
Units Peak area Peak area Peak area Peak area

MS:

MS ID:MS000276
Analysis ID:AN000327
Instrument Name:Agilent 6220 TOF
Instrument Type:TOF
MS Type:ESI
MS Comments:C18-PESI (+ve ESI) A 6220 ToF-MS (Agilent Technologies) was operated in both positive and negative electrospray ionization (ESI) modes using a scan range of 50?1,200 m/z. The mass accuracy and mass resolution were 5 parts per million (ppm) and 20,000 ppm, respectively. The instrument settings were as follows: nebulizer gas temperature 325°C, capillary voltage 3.5 kV, capillary temperature 300°C, fragmentor voltage 150 V, skimmer voltage 58 V, octapole voltage 250 V, cycle time 0.5 s, and run time 15.0 min.
Ion Mode:POSITIVE
  
MS ID:MS000277
Analysis ID:AN000328
Instrument Name:Agilent 6220 TOF
Instrument Type:TOF
MS Type:ESI
MS Comments:C18-NESI (-ve ESI) A 6220 ToF-MS (Agilent Technologies) was operated in both positive and negative electrospray ionization (ESI) modes using a scan range of 50?1,200 m/z. The mass accuracy and mass resolution were 5 parts per million (ppm) and 20,000 ppm, respectively. The instrument settings were as follows: nebulizer gas temperature 325°C, capillary voltage 3.5 kV, capillary temperature 300°C, fragmentor voltage 150 V, skimmer voltage 58 V, octapole voltage 250 V, cycle time 0.5 s, and run time 15.0 min.
Ion Mode:NEGATIVE
  
MS ID:MS000278
Analysis ID:AN000329
Instrument Name:Agilent 6220 TOF
Instrument Type:TOF
MS Type:ESI
MS Comments:HILIC-PESI (+ve ESI) A 6220 ToF-MS (Agilent Technologies) was operated in both positive and negative electrospray ionization (ESI) modes using a scan range of 50?1,200 m/z. The mass accuracy and mass resolution were 5 parts per million (ppm) and 20,000 ppm, respectively. The instrument settings were as follows: nebulizer gas temperature 325°C, capillary voltage 3.5 kV, capillary temperature 300°C, fragmentor voltage 150 V, skimmer voltage 58 V, octapole voltage 250 V, cycle time 0.5 s, and run time 15.0 min.
Ion Mode:POSITIVE
  
MS ID:MS000279
Analysis ID:AN000330
Instrument Name:Agilent 6220 TOF
Instrument Type:TOF
MS Type:ESI
MS Comments:HILIC-PESI (-ve ESI) A 6220 ToF-MS (Agilent Technologies) was operated in both positive and negative electrospray ionization (ESI) modes using a scan range of 50?1,200 m/z. The mass accuracy and mass resolution were 5 parts per million (ppm) and 20,000 ppm, respectively. The instrument settings were as follows: nebulizer gas temperature 325°C, capillary voltage 3.5 kV, capillary temperature 300°C, fragmentor voltage 150 V, skimmer voltage 58 V, octapole voltage 250 V, cycle time 0.5 s, and run time 15.0 min.
Ion Mode:NEGATIVE