Summary of Study ST001720

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001053. The data can be accessed directly via it's Project DOI: 10.21228/M8KQ59 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001720
Study TitleMetabolomics Analysis of ACTG Cohort -Update (part-II)
Study TypeMetabolomics Analysis
Study SummaryGlobal metabolomics analysis of ACTG cohort
Institute
The Wistar Institute
Last NameAbdel-Mohsen
First NameMohamed
Address3601 Spruce St, Philadelphia, PA 19104
Emailmmohsen@wistar.org
Phone215-898-6008
Submit Date2021-01-07
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2021-05-25
Release Version1
Mohamed Abdel-Mohsen Mohamed Abdel-Mohsen
https://dx.doi.org/10.21228/M8KQ59
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002803
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Vanquish
Column SeQuant ZIC-pHILIC (150 x 2.1mm,5um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF-X Orbitrap
Ion Mode UNSPECIFIED
Units Peak Area

MS:

MS ID:MS002598
Analysis ID:AN002803
Instrument Name:Thermo Q Exactive HF-X Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The following parameters were used for the MS analysis: sheath gas flow rate, 40; auxiliary gas flow rate, 10; sweep gas flow rate, 1; auxiliary gas heater temperature, 350 °C; spray voltage, 3.5 kV for positive mode and 3.2 kV for negative mode; capillary temperature, 325 °C; and funnel RF level, 40. All samples were analyzed by full MS with polarity switching. The QC sample was analyzed at the start of the sample sequence and after every 8-14 samples. The QC sample was also analyzed by data-dependent MS/MS with separate runs for positive and negative ion modes. Full MS scans were acquired at 120,000 resolution with a scan range of 65-975 m/z. Data-dependent MS/MS scans were acquired for the top 10 highest intensity ions at 15,000 resolution with an isolation width of 1.0 m/z and stepped normalized collision energy of 20-40-60. Data analysis was performed using Compound Discoverer 3.1 (ThermoFisher Scientific) with separate analyses for positive and negative polarities. Retention time alignment used the adaptative curve model with 0.3 min maximum shift, 5 ppm mass tolerance, and 3 S/N threshold. Peak detection required less than 5 ppm mass error for extracted ion chromatograms with a 50,000 minimum peak intensity. [M+H]+1 and [M-H]-1 were set as base ions with consideration for other adducts. Peaks were required to have a width at half height less than 0.5 min and a minimum of 5 scans. Components that had only a monoisotopic peak and no further isotopes were not considered. The maximum element count for isotope pattern modeling was C90H190N10Na2O15P3S5. Compounds were grouped across samples with 5 ppm mass error and 0.3 min retention time shift. Peaks not detected initially in a given sample were determined using the fill gaps algorithm with 5 ppm mass error and 1.5 S/N threshold with real peak detection. The gap function uses a priority system to determine missing values: 1) matching detected ions based on expected m/z and retention time regardless of adduct assignment, 2) re-detecting peaks at lower thresholds, 3) simulating peaks based on expected m/z, and 4) imputing spectrum noise based on detection limit values. Compound quantifications were corrected for instrument drift by QC areas using the cubic spline regression model. Each compound was required to be detected in at least 40% of QC runs with an RSD less than 50%. Metabolites were identified by accurate mass (5 ppm mass error) and retention time (0.5 min shift) using a database generated from pure standards or by accurate mass and MS2 spectra using the mzCloud spectral database (mzCloud.org), specifically the ‘Endogenous Metabolites’ and ‘Steroids/Vitamins/Hormones’ compound classes, and selecting the best matches with HighChem HighRes identity search match factors of 50 or greater. Results were manually processed to remove entries with apparent peak mis-integrations and correct commonly misannotated metabolites. Positive and negative data sets of identified compounds were merged, and the preferred polarity was selected for compounds identified in both polarities. Compound quantifications were normalized per volume plasma injected, which was equivalent for all samples. Values from the ACTG study were further normalized to the summed area of identified metabolites in each sample. For compounds identified multiple times at different retention times, a single entry was selected with priority given to standards database matches followed by greater mzCloud match factors and peak areas.
Ion Mode:UNSPECIFIED
  logo