Summary of Study ST001789

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001133. The data can be accessed directly via it's Project DOI: 10.21228/M87X24 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)
Study IDST001789
Study TitleAcute metabolomic changes of plasma in response to endurance exercise
Study SummaryThis study compared the metabolomic differences in plasma of young adult volunteers between before (Pre), immediately after the run (Time 0) and 60 minutes after the 90-min run (Time 60).
Institute
University of California, San Diego
DepartmentDepartment of Medicine
LaboratoryThe Naviaux Lab
Last NameNaviaux
First NameRobert
Address214 Dickinson St., Room C-107
Emailrnaviaux@health.ucsd.edu
Phone619-543-2904
Submit Date2021-04-21
Num Groups3
Total Subjects19
Num Males13
Num Females6
Analysis Type DetailLC-MS
Release Date2022-01-06
Release Version1
Robert Naviaux Robert Naviaux
https://dx.doi.org/10.21228/M87X24
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002901 AN002902
Analysis type MS MS
Chromatography type HILIC Reversed phase
Chromatography system Shimadzu 20AD Shimadzu 20AD
Column Shodex NH2P-40 2E (250 x 2mm,4um) Raptor Biphenyl (150 × 2.1 mm,2.7um)
MS Type ESI ESI
MS instrument type Triple quadrupole Triple quadrupole
MS instrument name ABI Sciex 5500 QTrap ABI Sciex 5500 QTrap
Ion Mode UNSPECIFIED UNSPECIFIED
Units peak area peak area

MS:

MS ID:MS002693
Analysis ID:AN002901
Instrument Name:ABI Sciex 5500 QTrap
Instrument Type:Triple quadrupole
MS Type:ESI
MS Comments:The MS/MS detection was performed using electrospray ionization (ESI) and advanced scheduled multiple reaction monitoring (MRM) with dynamic windowing and fast polarity switch. The ESI source conditions were set as follows: electrospray voltage of -4500 V for negative mode and 5500 V for positive mode, source temperature of 500 °C, curtain gas of 30, ion source gas 1, and gas 2 of 35 psi, respectively.Peaks were manually inspected and integrated using Multiquant 3.0.2. The peak area was exported to excel for further analysis.
Ion Mode:UNSPECIFIED
Collision Energy:Compound-dependent
Ion Source Temperature:500
Ion Spray Voltage:-4500 V for negative and 5500 V for positive
Ionization:ESI
Mass Accuracy:0.1 Da
  
MS ID:MS002694
Analysis ID:AN002902
Instrument Name:ABI Sciex 5500 QTrap
Instrument Type:Triple quadrupole
MS Type:ESI
MS Comments:The MS/MS detection was performed using electrospray ionization (ESI) and advanced scheduled multiple reaction monitoring (MRM) with dynamic windowing and fast polarity switch. The ESI source conditions were set as follows: electrospray voltage of -4500 V for negative mode and 5500 V for positive mode, source temperature of 500 °C, curtain gas of 30, ion source gas 1, and gas 2 of 35 psi, respectively.Peaks were manually inspected and integrated using Multiquant 3.0.2. The peak area was exported to excel for further analysis.
Ion Mode:UNSPECIFIED
Collision Energy:Compound-dependent
Ion Source Temperature:500
Ion Spray Voltage:-4500 V for negative and 5500 V for positive
Ionization:ESI
Mass Accuracy:0.1 Da
  logo