Summary of Study ST003246

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR002016. The data can be accessed directly via it's Project DOI: 10.21228/M81C00 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST003246
Study TitleEffects of mitoregulin loss on cardiac and mitochondrial lipids in aged male mice
Study SummaryCardiac lipidome analysis in aged (21 to 23-months old) male wildtype and mtln knockout mice
Institute
University of Iowa
Last NameBoudreau
First NameRyan
Address4334 PBDB, 169 Newton Rd, Iowa City, IA 52242
Emailryan-boudreau@uiowa.edu
Phone3193535573
Submit Date2024-06-02
Raw Data AvailableYes
Raw Data File Type(s)mzML
Analysis Type DetailLC-MS
Release Date2024-12-31
Release Version1
Ryan Boudreau Ryan Boudreau
https://dx.doi.org/10.21228/M81C00
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN005316 AN005317
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Agilent 1290 Infinity Agilent 1290 Infinity
Column Waters ACQUITY UPLC CSH C18 (100 x 2.1mm,1.7um) Waters ACQUITY UPLC CSH C18 (100 x 2.1mm,1.7um)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name Agilent 6545 QTOF Agilent 6545 QTOF
Ion Mode POSITIVE NEGATIVE
Units counts normalized to total lipid signal counts normalized to total lipid signal

MS:

MS ID:MS005046
Analysis ID:AN005316
Instrument Name:Agilent 6545 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:For positive mode, the source gas temperature was set to 225 °C, with a drying gas flow of 11 L/minute, nebulizer pressure of 40 psig, sheath gas temp of 350 °C and sheath gas flow of 11 L/minute. VCap voltage is set at 3500 V, nozzle voltage 500V, fragmentor at 110 V, skimmer at 85 V and octopole RF peak at 750 V. For data processing, Agilent MassHunter (MH) Workstation and software packages MH Qualitiative and MH Quantitative were used. The pooled QC (n=8) and process blank (n=4) were injected throughout the sample queue to ensure the reliability of acquired lipidomics data. For lipid annotation, accurate mass and MS/MS matching was used with the Agilent Lipid Annotator library. Results from the positive and negative ionization modes from Lipid Annotator were merged based on the class of lipid identified. Data exported from MH Quantitative was evaluated using Excel where initial lipid targets are parsed based on the following criteria. Only lipids with relative standard deviations (RSD) less than 30% in QC samples are used for data analysis. Additionally, only lipids with background AUC counts in process blanks that are less than 30% of QC are used for data analysis. The parsed excel data tables are normalized based on the ratio to class-specific internal standards, then to sum prior to statistical analysis.
Ion Mode:POSITIVE
  
MS ID:MS005047
Analysis ID:AN005317
Instrument Name:Agilent 6545 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:For negative mode, the source gas temperature was set to 300 °C, with a drying gas flow of 11 L/minute, a nebulizer pressure of 30 psig, sheath gas temp of 350 °C and sheath gas flow 11 L/minute. VCap voltage was set at 3500 V, nozzle voltage 75 V, fragmentor at 175 V, skimmer at 75 V and octopole RF peak at 750 V. For data processing, Agilent MassHunter (MH) Workstation and software packages MH Qualitiative and MH Quantitative were used. The pooled QC (n=8) and process blank (n=4) were injected throughout the sample queue to ensure the reliability of acquired lipidomics data. For lipid annotation, accurate mass and MS/MS matching was used with the Agilent Lipid Annotator library. Results from the positive and negative ionization modes from Lipid Annotator were merged based on the class of lipid identified. Data exported from MH Quantitative was evaluated using Excel where initial lipid targets are parsed based on the following criteria. Only lipids with relative standard deviations (RSD) less than 30% in QC samples are used for data analysis. Additionally, only lipids with background AUC counts in process blanks that are less than 30% of QC are used for data analysis. The parsed excel data tables are normalized based on the ratio to class-specific internal standards, then to sum prior to statistical analysis.
Ion Mode:NEGATIVE