                      INSTRUMENT CONTROL PARAMETERS:    7890A
                      ---------------------------------------

   C:\MSDCHEM\1\5975\PROFILING HP-5MS.M
      Sat Aug 28 17:58:47 2010

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  GC ALS
Mass Spectrometer :  Enabled

Oven
Equilibration Time                           0.25 min
Oven Program                                 On
    70 C for 5 min
    then 5 C/min to 310 C for 20 min
Run Time                                     73 min





Front Injector
Syringe Size                                 10 L
Injection Volume                             3 L
Injection Repetitions                        1
Injection Delay                              0 sec
Solvent A Washes (PreInj)                    7
Solvent A Washes (PostInj)                   7
Solvent A Volume                             8 L
Solvent B Washes (PreInj)                    0
Solvent B Washes (PostInj)                   0
Solvent B Volume                             8 L
Sample Washes                                1
Sample Wash Volume                           8 L
Sample Pumps                                 5
Dwell Time (PreInj)                          0 min
Dwell Time (PostInj)                         0 min
Solvent Wash Draw Speed                      300 L/min
Solvent Wash Dispense Speed                  6000 L/min
Sample Wash Draw Speed                       300 L/min
Sample Wash Dispense Speed                   6000 L/min
Injection Dispense Speed                     6000 L/min
Viscosity Delay                              0 sec
Sample Depth                                 Disabled


Sample Overlap
Sample overlap is not enabled



Front SS Inlet He
Mode                                         Split
Heater                                       On    250 C
Pressure                                     On    26.015 psi
Total Flow                                   On    19.5 mL/min
Septum Purge Flow                            On    3 mL/min
Gas Saver                                    Off
Split Ratio                                  10 :1
Split Flow                                   15 mL/min


Thermal Aux 2 {MSD Transfer Line}
Heater                                       On
Temperature Program                          On
    250 C for 0 min
Run Time                                     73 min


Column #1
HP-5  5% Phenyl Methyl Siloxan: 958.50561
HP-5  5% Phenyl Methyl Siloxan
325 C: 60 m x 250 m x 0.25 m
In: Front SS Inlet He
Out: Vacuum

(Initial)                                    70 C
Pressure                                     26.015 psi
Flow                                         1.5 mL/min
Average Velocity                             31.867 cm/sec
Holdup Time                                  3.1381 min
Flow Program                                 On
    1.5 mL/min for 0 min
Run Time                                     73 min





Front Detector FID
Heater                                       Off
H2 Flow                                      Off
Air Flow                                     Off
Makeup Flow                                  Off
Const Col + Makeup                           Off
Flame                                        Off
Electrometer                                 Off









Aux EPC 1 He
Aux EPC 2 He
Aux EPC 3 He



Signals
 None
 Test Plot                                   Save Off
                                             50 Hz
 Test Plot                                   Save Off
                                             50 Hz
 Test Plot                                   Save Off
                                             50 Hz


                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : EI_atune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 14.20 min

EMV Mode                 : Gain Factor
Gain Factor              : 2.00
Resulting EM Voltage     : 1471

[Scan Parameters]

Low Mass                 : 50.0
High Mass                : 800.0
Threshold                : 100
Sample #                 : 3       A/D Samples    8
Plot 2 low mass          : 271.5
Plot 2 high mass         : 272.5

[MSZones]

MS Source                : 230 C   maximum 250 C
MS Quad                  : 150 C   maximum 200 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US71245796
                        -----------------------------

 Trace Ion Detection is ON.

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      28.950
 IONFOCUS    :      90.157
 ENTRANCE_LE :      12.500
 EMVOLTS     :    1435.294
                               Actual EMV  :    1470.59
                               GAIN FACTOR :       2.00
 AMUGAIN     :    1588.000
 AMUOFFSET   :     124.688
 FILAMENT    :       1.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      20.329
 MASSGAIN    :    -865.000   
 MASSOFFSET  :     -36.000   

                           END OF TUNE PARAMETERS
                      ------------------------------------



                       END OF INSTRUMENT CONTROL PARAMETERS
                       ------------------------------------