Return to study ST002701 main page

MB Sample ID: SA270920

Local Sample ID:CRS9
Subject ID:SU002806
Subject Type:Mammal
Subject Species:Mus musculus
Taxonomy ID:10090
Gender:Female

Select appropriate tab below to view additional metadata details:


Subject:

Subject ID:SU002806
Subject Type:Mammal
Subject Species:Mus musculus
Taxonomy ID:10090
Gender:Female

Factors:

Local Sample IDMB Sample IDFactor Level IDLevel ValueFactor Name
CRS9SA270920FL035198stressTreatment

Collection:

Collection ID:CO002799
Collection Summary:To precipitate proteins, we combined 100 μl of SERUM with 800 μl of methanol, incubated at -20 for 30 min, then centrifuged at 18000 rpm for 10 min. Subsequently, 850 μl of supernatant was transferred to a new tube and evaporated to dryness in a Thermo Savant SpeedVac. Then 200 μl of resuspension solvent methanol was added, compounds were dissolved by vortexing for 5 min and samples were clarified by centrifuged at 18000 rpm for 10 min.
Sample Type:Blood (serum)

Treatment:

Treatment ID:TR002815
Treatment Summary:Chronic restraint stress (RS) in mice were placed inside a plastic mouse holder for 4 h daily, from 10:00 am to 14:00 pm, and repeated for a total of 14 consecutive days. During non-stress periods, all mice were conventionally housed and allowed free access to food and water.

Sample Preparation:

Sampleprep ID:SP002812
Sampleprep Summary:To precipitate proteins, we combined 200 μl SERUM with 800 μl of methanol, incubated at -20 for 30 min, then centrifuged at 18000 rpm for 10 min. Subsequently, 850 μl of supernatant was transferred to a new tube and evaporated to dryness in a Thermo Savant SpeedVac. Then 200 μl of resuspension solvent methanol was added, compounds were dissolved by vortexing for 5 min and samples were clarified by centrifuged at 18000 rpm for 10 min.
Processing Storage Conditions:-80℃
Extract Storage:-20℃

Combined analysis:

Analysis ID AN004378
Analysis type MS
Chromatography type Reversed phase
Chromatography system Agilent 6546 Q-TOF
Column Waters Atlantis 3μm particle size T3 column (2.1 × 10 cm)
MS Type ESI
MS instrument type QTOF
MS instrument name Agilent 6545 QTOF
Ion Mode NEGATIVE
Units Abundance

Chromatography:

Chromatography ID:CH003283
Instrument Name:Agilent 6546 Q-TOF
Column Name:Waters Atlantis 3μm particle size T3 column (2.1 × 10 cm)
Column Temperature:40
Flow Gradient:A linear gradient of 0–95% eluent B over 10min. Eluent B was held at 100% for 2.5 min, then brought to 5% over 0.5 min and the column was re-equilibrated at 0% eluent B for 4 min
Flow Rate:0.25ml/min
Solvent A:100% water; 0.05% formic acid; 0.1% ammonium formate
Solvent B:100% acetonitrile
Chromatography Type:Reversed phase

MS:

MS ID:MS004127
Analysis ID:AN004378
Instrument Name:Agilent 6545 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:Fast data-dependent acquisition (DDA) MS/MS experiments were performed. The Progenesis QI (Nonlinear Dynamics, Newcastle, UK) was used for peak picking and alignment to screen the metabolic biomarkers that displayed significant changes between the control and the fructose-treated group. Molecular identification of the assigned biomarkers was accomplished by matching the acquired precursors and fragment ions against several standard metabolome databases including the Human Metabolome Database (http://www.hmdb.ca/), MassBank (http://www.massbank.jp/index.html), and METLIN (http://metlin.scripps.edu/index.php). Partial metabolite identification was further confirmed by comparison with available standards. Metabolic pathway enrichment analysis of these identified metabolic biomarkers was carried out by MetaboAnalyst 3.0
Ion Mode:NEGATIVE