MB Sample ID: SA291704

Local Sample ID:BPL1415 0689
Subject ID:SU002871
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Subject:

Subject ID:SU002871
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Factors:

Local Sample IDMB Sample IDFactor Level IDLevel ValueFactor Name
BPL1415 0689SA291704FL035828Caseliver_cancer_outcome

Collection:

Collection ID:CO002864
Collection Summary:Baseline visit blood samples were collected from ATBC participants after an overnight fast (usually at least 12 hours). Blood samples were centrifuged and serum was aliquoted and stored at -70C.
Sample Type:Blood (serum)

Treatment:

Treatment ID:TR002880
Treatment Summary:This is an nested case control study within the ATBC cohort. There were not treatment groups.

Sample Preparation:

Sampleprep ID:SP002877
Sampleprep Summary:The lipidome analysis was performed at the West Coast Metabolomics Center (University of California Davis Genome Center), using 50μL of serum as previously described in https://www.nature.com/articles/sdata2018263

Combined analysis:

Analysis ID AN004498 AN004499
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Agilent 1290 Infinity Agilent 1290 Infinity
Column Waters ACQUITY UPLC BEH C18 (100 x 2.1mm,1.7um) Waters ACQUITY UPLC BEH C18 (100 x 2.1mm,1.7um)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name Agilent 6550 QTOF Agilent 6530 QTOF
Ion Mode NEGATIVE POSITIVE
Units normalized peak intensity normalized peak intensity

Chromatography:

Chromatography ID:CH003380
Chromatography Summary:The LC/QTOFMS analyses are performed using an Agilent 1290 Infinity LC system (G4220A binary pump, G4226A autosampler, and G1316C Column Thermostat) coupled to either an Agilent 6530 (positive ion mode) or an Agilent 6550 mass spectrometer equipped with an ion funnel (iFunnel) (negative ion mode). Lipids are separated on an Acquity UPLC CSH C18 column (100 x 2.1 mm; 1.7 µm) maintained at 65°C at a flow-rate of 0.6 mL/min. Solvent pre-heating (Agilent G1316) was used. The mobile phases consist of 60:40 acetonitrile:water with 10 mM ammonium formate and 0.1% formic acid (A) and 90:10 propan-2-ol:acetonitrile with 10 mM ammonium formate and 0.1% formic acid. The gradient is as follows: 0 min 85% (A); 0–2 min 70% (A); 2–2.5 min 52% (A); 2.5–11 min 18% (A); 11–11.5 min 1% (A); 11.5–12 min 1% (A); 12–12.1 min 85% (A); 12.1–15 min 85% (A). A sample volume of 3 µL is used for the injection. Sample temperature is maintained at 4°C in the autosampler.
Instrument Name:Agilent 1290 Infinity
Column Name:Waters ACQUITY UPLC BEH C18 (100 x 2.1mm,1.7um)
Column Temperature:65
Flow Gradient:0 min 85% (A); 0-2 min 70% (A); 2-2.5 min 52% (A); 2.5-11 min 18% (A); 11-11.5 min 1% (A); 11.5-12 min 1% (A); 12-12.1 min 85% (A); 12.1-15 min 85% (A).
Flow Rate:0.6
Solvent A:60% acetonitrile/40% water; 0.1% formic acid; 10 mM ammonium formate
Solvent B:90% isopropanol/10% acetonitrile; 0.1% formic acid; 10 mM ammonium formate
Chromatography Type:Reversed phase

MS:

MS ID:MS004245
Analysis ID:AN004498
Instrument Name:Agilent 6550 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:The quadrupole/time-of-flight (QTOF) mass spectrometers are operated with electrospray ionization (ESI) performing full scan in the mass range m/z 65–1700 in positive (Agilent 6530, equipped with a JetStreamSource) and negative (Agilent 6550, equipped with a dual JetStream Source) modes producing both unique and complementary spectra. Instrument parameters are as follows (positive mode) Gas Temp 325°C, Gas Flow 8 l/min, Nebulizer 35 psig, Sheath Gas 350°C, Sheath Gas Flow 11, Capillary Voltage 3500 V, Nozzle Voltage 1000V, Fragmentor 120V, Skimmer 65V. Data (both profile and centroid) are collected at a rate of 2 scans per second. In negative ion mode, Gas Temp 200°C, Gas Flow 14 l/min, Fragmentor 175V, with the other parameters identical to positive ion mode. For the 6530 QTOF, a reference solution generating ions of 121.050 and 922.007 m/z in positive mode and 119.036 and 966.0007 m/z in negative mode, and these are used for continuous mass correction. For the 6550, the reference solution is introduced via a dual spray ESI, with the same ions and continuous mass correction. Samples are injected (1.7 μl in positive mode and 5 μl in negative ion mode) with a needle wash for 20 seconds (wash solvent is isopropanol). The valve is switched back and forth during the run for washing; this has been shown to be essential for reducing carryover of less polar lipids.
Ion Mode:NEGATIVE
  
MS ID:MS004246
Analysis ID:AN004499
Instrument Name:Agilent 6530 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:The quadrupole/time-of-flight (QTOF) mass spectrometers are operated with electrospray ionization (ESI) performing full scan in the mass range m/z 65–1700 in positive (Agilent 6530, equipped with a JetStreamSource) and negative (Agilent 6550, equipped with a dual JetStream Source) modes producing both unique and complementary spectra. Instrument parameters are as follows (positive mode) Gas Temp 325°C, Gas Flow 8 l/min, Nebulizer 35 psig, Sheath Gas 350°C, Sheath Gas Flow 11, Capillary Voltage 3500 V, Nozzle Voltage 1000V, Fragmentor 120V, Skimmer 65V. Data (both profile and centroid) are collected at a rate of 2 scans per second. In negative ion mode, Gas Temp 200°C, Gas Flow 14 l/min, Fragmentor 175V, with the other parameters identical to positive ion mode. For the 6530 QTOF, a reference solution generating ions of 121.050 and 922.007 m/z in positive mode and 119.036 and 966.0007 m/z in negative mode, and these are used for continuous mass correction. For the 6550, the reference solution is introduced via a dual spray ESI, with the same ions and continuous mass correction. Samples are injected (1.7 μl in positive mode and 5 μl in negative ion mode) with a needle wash for 20 seconds (wash solvent is isopropanol). The valve is switched back and forth during the run for washing; this has been shown to be essential for reducing carryover of less polar lipids.
Ion Mode:POSITIVE