Return to study ST000017 main page
MB Sample ID: SA001065
Local Sample ID: | S00009499 |
Subject ID: | SU000033 |
Subject Type: | Animal |
Subject Species: | Rattus norvegicus |
Taxonomy ID: | 10116 |
Genotype Strain: | Low Capacity Running | High Capacity Running |
Species Group: | Mammal |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN000034 | AN000035 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | Reversed phase | Reversed phase |
Chromatography system | Agilent 1200 | Agilent 1200 |
Column | Waters Acquity HSS T3 reversed-phase | Waters Acquity HSS T3 reversed-phase |
MS Type | ESI | ESI |
MS instrument type | QTOF | QTOF |
MS instrument name | Agilent 6530 QTOF | Agilent 6530 QTOF |
Ion Mode | POSITIVE | NEGATIVE |
Units | peak area | peak area |
Chromatography:
Chromatography ID: | CH000017 |
Chromatography Summary: | Positive ion detection |
Methods Filename: | EX00166-LCMS-method.pdf |
Instrument Name: | Agilent 1200 |
Column Name: | Waters Acquity HSS T3 reversed-phase |
Column Temperature: | 40C |
Flow Gradient: | 0-0.5 min 1% B, 0.5-2 min 1-99% B, 2-6 min 99% B, 6-9 min 1% B. |
Flow Rate: | 0.35 mL/min |
Solvent A: | 100% water; 0.1% formic acid |
Solvent B: | 100% methanol; 0.1% formic acid |
Chromatography Type: | Reversed phase |
Chromatography ID: | CH000018 |
Chromatography Summary: | Negative ion detection |
Methods Filename: | EX00166-LCMS-method.pdf |
Instrument Name: | Agilent 1200 |
Column Name: | Waters Acquity HSS T3 reversed-phase |
Column Temperature: | 40C |
Flow Gradient: | 0-0.5 min 1% B, 0.5-2 min 1-99% B, 2-6 min 99% B, 6-9 min 1% B. |
Flow Rate: | 0.35 mL/min |
Solvent A: | 100% water; 0.1% ammonium bicarbonate |
Solvent B: | 100% methanol; 0.1% ammonium bicarbonate |
Chromatography Type: | Reversed phase |