Combined analysis:

Analysis ID AN002391 AN002392
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Thermo Dionex Ultimate 3000 Thermo Dionex Ultimate 3000
Column Agilent Zorbax Eclipse Plus C18 (100 x 2.1mm, 1.8 um) Agilent Zorbax Eclipse Plus C18 (100 x 2.1mm, 1.8 um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Plus Orbitrap Thermo Q Exactive Plus Orbitrap
Ion Mode POSITIVE NEGATIVE
Units peak area peak area

Chromatography:

Chromatography ID:CH001758
Chromatography Summary:Chromatographic conditions RPLC separation was performed using Zorbax SB columns (2.1 X 50mm, 1.8 Micron, 600 Bar; 827700-914) purchased from Agilent Technologies (Santa Clara, CA, USA). Mobile phases for RPLC consisted of 0.06% acetic acid in water (phase A) and 0.06% acetic acid in MeOH (phase B). Metabolites were eluted from the column at a flow rate of 0.6 mL/min, leading to a backpressure of 220– 280 bar at 99% phase A. A linear 1%–80% phase B gradient was applied over 9–10 min. The oven temperature was set to 60C, and the sample injection volume was 5 mL.
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:Agilent Zorbax Eclipse Plus C18 (100 x 2.1mm, 1.8 um)
Chromatography Type:Reversed phase