Return to study ST001735 main page
MB Sample ID: SA162751
Local Sample ID: | CUcovID_0061 |
Subject ID: | SU001812 |
Subject Type: | Human |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN002824 | AN002825 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | Reversed phase | Reversed phase |
Chromatography system | Thermo Vanquish | Thermo Vanquish |
Column | Phenomenex Kinetex C18 (150 x 2.1mm,1.7um) | Phenomenex Kinetex C18 (150 x 2.1mm,1.7um) |
MS Type | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap | Thermo Q Exactive Orbitrap |
Ion Mode | POSITIVE | NEGATIVE |
Units | peak area | peak area |
Chromatography:
Chromatography ID: | CH002088 |
Chromatography Summary: | After sample randomization, 10 μL of extracts were injected into a Thermo Vanquish UHPLC system (San Jose, CA, USA) and resolved on a Kinetex C18 column (150 × 2.1 mm, 1.7 μm, Phenomenex, Torrance, CA, USA) at 450 μL/min through a 5 min gradient from 5 to 95% organic solvent B (mobile phases: A = water, 0.1% formic acid; B = acetonitrile, 0.1% formic acid) in positive ion mode. Solvent gradient: 0-0.5 min 5% B, 0.5-1.1 min 5-95% B, 1.1-2.75 min hold at 95% B, 2.75-3 min 95-5% B, 3-5 min hold at 5% B. |
Instrument Name: | Thermo Vanquish |
Column Name: | Phenomenex Kinetex C18 (150 x 2.1mm,1.7um) |
Column Temperature: | 45 |
Flow Gradient: | Solvent gradient: 0-0.5 min 5% B, 0.5-1.1 min 5-95% B, 1.1-2.75 min hold at 95% B, 2.75-3 min 95-5% B, 3-5 min hold at 5% B |
Flow Rate: | 450 ul/min |
Solvent A: | 100% water; 0.1% formic acid |
Solvent B: | 100% acetonitrile; 0.1% formic acid |
Chromatography Type: | Reversed phase |
Chromatography ID: | CH002089 |
Chromatography Summary: | After sample randomization, 10 μL of extracts were injected into a Thermo Vanquish UHPLC system (San Jose, CA, USA) and resolved on a Kinetex C18 column (150 × 2.1 mm, 1.7 μm, Phenomenex, Torrance, CA, USA) at 450 μL/min through a 5 min gradient from 0 to 100% phase B (phase A was 5% acetonitrile, 95% water, 1 mM ammonium acetate, phase B was 95% acetonitrile, 5% water, 0.5 mM ammonium acetate). Solvent gradient: 0-0.5 min 0% B; 0.5-1.1 min 0-100% B, 1.1-2.75 min hold at 100% B, 2.75-3 min 100-0% B, 3-5 min hold at 0% B. |
Instrument Name: | Thermo Vanquish |
Column Name: | Phenomenex Kinetex C18 (150 x 2.1mm,1.7um) |
Column Temperature: | 45 |
Flow Gradient: | Solvent gradient: 0-0.5 min 0% B; 0.5-1.1 min 0-100% B, 1.1-2.75 min hold at 100% B, 2.75-3 min 100-0% B, 3-5 min hold at 0% B. |
Flow Rate: | 450 uL/min |
Solvent A: | 5% acetonitrile/95% water; 1 mM ammonium acetate |
Solvent B: | 95% acetonitrile/5% water; 0.5 mM ammonium acetate |
Chromatography Type: | Reversed phase |