Return to study ST001851 main page

MB Sample ID: SA172899

Local Sample ID:KWS10653
Subject ID:SU001928
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002999
Analysis type MS
Chromatography type Reversed phase
Chromatography system Exion LC
Column InertSustain C18
MS Type ESI
MS instrument type Triple quadrupole
MS instrument name ABI Sciex Triple Quad 6500+
Ion Mode UNSPECIFIED
Units nM

Chromatography:

Chromatography ID:CH002224
Chromatography Summary:2 μL was injected to LC/ESI-MS/MS (Triple Quad 6500+ tandem mass spectrometer, equipped with an ESI probe and Exion LC AD ultra-high pressure liquid chromatography system; SCIEX). A separation column, InertSustain C18 (150 mm × 2.1 mm ID, 3 μm particle size; GL Sciences Inc.), was utilized at 40 °C. A mixture of 10 mM ammonium acetate and acetonitrile was used as the eluent and the separation was carried out by linear gradient elution at a flow rate of 0.2 mL/min. The mobile phase composition was gradually changed as follows: ammonium acetate-acetonitrile (86:14, v/v) for 0.5 min, (78:22, v/v) for 0.5-5 min, (72:28, v/v) for 5-28 min, (46:54, v/v) for 28-55 min, (2:98, v/v) for 55-66 min, and (2:98, v/v) for 4 min. The total run time was 70 min.
Instrument Name:Exion LC
Column Name:InertSustain C18
Column Temperature:40
Flow Gradient:ammonium acetate-acetonitrile (86:14, v/v) for 0.5 min, (78:22, v/v) for 0.5-5 min, (72:28, v/v) for 5-28 min, (46:54, v/v) for 28-55 min, (2:98, v/v) for 55-66 min, and (2:98, v/v) for 4 min. The total run time was 70 min.
Flow Rate: 0.2 mL/min
Solvent A:100% water; 10mM ammonium acetate
Solvent B:100% acetonitrile
Chromatography Type:Reversed phase
  logo