Return to study ST001991 main page

MB Sample ID: SA186035

Local Sample ID:5004195
Subject ID:SU002072
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003248
Analysis type MS
Chromatography type Reversed phase
Chromatography system Waters
Column Waters ACQUITY UPLC HSS T3 (100 x 2.1mm,1.8um)
MS Type ESI
MS instrument type Triple quadrupole
MS instrument name ABI Sciex 5500 QTrap
Ion Mode NEGATIVE
Units ng/ml

Chromatography:

Chromatography ID:CH002393
Chromatography Summary:The analyses were performed on an ACQUITY HSS T3 (2.1×100 mm, 1.8 μm) column, Waters (Milford), coupled to a triple quadrupole mass spectrometer (Waters Corporation, Milford, USA) with an atmospheric electrospray interface operating in negative ion mode. Separation was performed using gradient elution with 0.1 % formic acid in water (v/v) (A) and 0.1 % formic acid in acetonitrile:methanol (3:1, v/v) (B) at a flow rate of 0.5 mL/ min. Gradient program was 0 min 15 % B, 0-1 min; 30 % B, 1- 16 min; 16-18 min; 70 % B, 18–23 min 100 % B, and equilibrium time between runs was 7 min. The injection volume was 5 μL and the column was kept at 35 °C.
Instrument Name:Waters
Column Name:Waters ACQUITY UPLC HSS T3 (100 x 2.1mm,1.8um)
Column Temperature:35
Flow Gradient:0 min 15 % B, 0-1 min; 30 % B, 1- 16 min; 16-18 min; 70 % B, 18–23 min 100 % B, and equilibrium time between runs was 7 min.
Flow Rate:0.5 mL/min
Solvent A:100% water; 0.1% formic acid
Solvent B:75% acetonitrile/25% methanol; 0.1% formic acid
Chromatography Type:Reversed phase