MB Sample ID: SA187840

Local Sample ID:Veh-2
Subject ID:SU002088
Subject Type:Cultured cells
Subject Species:Mus musculus
Taxonomy ID:10090

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003270
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Dionex Ultimate 3000
Column Waters XSelect HSS C18 (250 x 4.6mm)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Exactive Plus Orbitrap
Ion Mode NEGATIVE
Units peak area

Chromatography:

Chromatography ID:CH002414
Chromatography Summary:The sample run order was randomized to reduce bias from instrument drift. 10 µl sample was subjected to LC-MS analysis on a Exactive Plus Orbitrap mass spectrometer (Thermo Fisher Scientific) coupled to an Ultimate 3000 UHPLC system (Thermo Fisher Scientific). Reversed-phase chromatography mode was used to separate compounds on a Xselect C18 HSS column (Waters) with solvent A (97:3 water:methanol (v/v), 10 mM tributylamine, and 15 mM acetic acid ) and solvent B (methanol). The flow rate was 200 µl/min, and the total run time was 25 min. The gradient was 0 min, 0% B; 5 min, 20% B; 7.5 min, 55% B; 15 min, 65% B; 17.5 min, 95% B; and 21 min, 0% B. The mass spectrometer was operated in a negative-ion mode at a resolution of 140,000 at m/z 200 and with a scan range of 85 to 1000 m/z.
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:Waters XSelect HSS C18 (250 x 4.6mm)
Flow Gradient:The gradient was 0 min, 0% B; 5 min, 20% B; 7.5 min, 55% B; 15 min, 65% B; 17.5 min, 95% B; and 21 min, 0% B.
Flow Rate:200 ul/min
Solvent A:97% water/3% methanol; 15 mM acetic acid; 10 mM tributylamine
Solvent B:100% methanol
Chromatography Type:Reversed phase