Return to study ST002056 main page

MB Sample ID: SA193793

Local Sample ID:CarnsTG2
Subject ID:SU002138
Subject Type:Mammal
Subject Species:Mus musculus
Age Or Age Range:8-12 weeks
Gender:Male

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003348
Analysis type MS
Chromatography type GC
Chromatography system Agilent 6890N
Column Agilent DB5-MS (60m × 0.25mm, 0.25um)
MS Type EI
MS instrument type GC x GC-TOF
MS instrument name Leco Pegasus III GC TOF
Ion Mode UNSPECIFIED
Units intensity

Chromatography:

Chromatography ID:CH002479
Chromatography Summary:The extracted and derivatized samples were analyzed on a LECO Pegasus GC×GC-TOF MS instrument (LECO Corp., St. Joseph, MI, USA) coupled to an Agilent 6890 gas chromatography and a Gerstel MPS2 autosampler (GERSTEL Inc., Linthicum, MD, USA), featuring a LECO two-stage cryogenic modulator and secondary oven. The primary column was a 60 m × 0.25 mm 1dc × 0.25 µm 1df DB-5 ms GC capillary column (phenyl arylene polymer virtually equivalent to (5%-phenyl)-methylpolysiloxane). The secondary GC column 1 m × 0.25 mm 2dc × 0.25 µm 2df, DB-17 ms ((50% phenyl)-methylpolysiloxane) was placed inside the secondary GC oven following the thermal modulator. Both columns were obtained from Agilent Technologies (Agilent Technologies J&W, Santa Clara, CA, USA) and were connected through a press fit connector. The helium carrier gas (99.999% purity) flow rate was set to 2.0 mL/min at a corrected constant flow via pressure ramps. The inlet temperature was set at 280 °C. The primary column temperature was programmed with an initial temperature of 60 °C for 0.5 min, then ramped at 5°C/min to 270 °C, and maintained for 15 min. The secondary column temperature program was set to an initial temperature of 70 °C for 0.5 min and then ramped at the same temperature gradient employed in the first column to 280 °C, accordingly. The thermal modulator was set to 15 °C relative to the primary oven and a modulation time was 2s. The mass range was set as 29-800 m/z with an acquisition rate of 200 mass spectra per second. The ion source chamber was 230°C with the transfer line temperature of 280°C, and the detector voltage was 1440 V with electron energy of 70 eV. The acceleration voltage was turned on after a solvent delay of 544 s, the split ratio was set at 10:1.
Instrument Name:Agilent 6890N
Column Name:Agilent DB5-MS (60m × 0.25mm, 0.25um)
Chromatography Type:GC
  logo