Return to study ST002061 main page

MB Sample ID: SA194026

Local Sample ID:WT3M2
Subject ID:SU002143
Subject Type:Cultured cells
Subject Species:Mus musculus

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003360
Analysis type MS
Chromatography type HILIC
Chromatography system ThermoFisher Ultimate 3000 UHPLC system
Column Waters BEH Amide
MS Type ESI
MS instrument type Hybrid Quadrupole-Orbitrap™
MS instrument name ThermoFisher Q Exactive™ Hybrid Quadrupole-Orbitrap™ Mass Spectrometry (QE)
Ion Mode NEGATIVE
Units pmoles/I

Chromatography:

Chromatography ID:CH002486
Chromatography Summary:Chromatographic separation was performed on a ThermoFisher Ultimate 3000 UHPLC system with a Waters BEH Amide column (2.1mm × 100 mm, 1.7 μm). The injection volume was 2μL and the flow rate was 0.25 mL/min. The column temperature was 15°C. The mobile phases consisted of water with 0.01% formic acid and 2 mM ammonium formate (phase A) and acetonitrile (phase B). A linear gradient elution was performed with the following program: 0 min, 90%B; 4 min, 85% B; 11 min, 75%B; 14 min, 70%B, 14.5min, 50%B and held to 17 min; 17.1 min, 90%B and held to 20.01 min.
Instrument Name:ThermoFisher Ultimate 3000 UHPLC system
Column Name:Waters BEH Amide
Column Temperature:15
Flow Gradient:0 min, 90%B; 4 min, 85% B; 11 min, 75%B; 14 min, 70%B, 14.5min, 50%B and held to 17 min; 17.1 min, 90%B and held to 20.01 min.
Flow Rate:0.25ml/min
Solvent A:100% water; 0.01% formic acid; 2 mM ammonium formate
Solvent B:100% acetonitrile
Chromatography Type:HILIC