Return to study ST002330 main page
MB Sample ID: SA229721
Local Sample ID: | HILIC-RN19-neg |
Subject ID: | SU002417 |
Subject Type: | Plant |
Subject Species: | Brassica napus |
Taxonomy ID: | 3708 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN003801 | AN003802 | AN003803 |
---|---|---|---|
Analysis type | MS | MS | MS |
Chromatography type | HILIC | HILIC | Reversed phase |
Chromatography system | Thermo Dionex Ultimate 3000 RS | Thermo Dionex Ultimate 3000 RS | Thermo Dionex Ultimate 3000 RS |
Column | Agilent InfinityLab Poroshell 120 HILIC-Z (100 x 2.1mm,2.7um) | Agilent InfinityLab Poroshell 120 HILIC-Z (100 x 2.1mm,2.7um) | Phenomonex Kinetex C18 (50 x 2.1mm,2.6um) |
MS Type | ESI | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap | Thermo Q Exactive Orbitrap | Thermo Q Exactive Orbitrap |
Ion Mode | POSITIVE | NEGATIVE | NEGATIVE |
Units | normalized relative level | normalized relative level | normalized relative level |
Chromatography:
Chromatography ID: | CH002812 |
Chromatography Summary: | Primary metabolites were resolved with an InfinityLab Poroshell 120 HILIC-Z column (2.1 × 100 mm, 2.7 µm; Agilent, Santa Clara, CA, US), using a flow rate of 0.2 mL min-1 with a mobile phase of 10 mM ammonium bicarbonate in water (A) and 10 mM ammonium bicarbonate in 95% acetonitrile (B). The following gradient was used to elute the analytes: mobile phase 100% B decreased to 90% over 2.5 min and to 50% over the next 5 min and returned to 100% over 0.5 min for 12 min of column re-equilibration. |
Instrument Name: | Thermo Dionex Ultimate 3000 RS |
Column Name: | Agilent InfinityLab Poroshell 120 HILIC-Z (100 x 2.1mm,2.7um) |
Column Temperature: | RT |
Flow Rate: | 0.2 ml/min |
Injection Temperature: | 4 |
Solvent A: | 100% water; 10 mM ammonium bicarbonate |
Solvent B: | 95% acetonitrile/5% water; 10 mM ammonium bicarbonate |
Chromatography Type: | HILIC |
Chromatography ID: | CH002813 |
Chromatography Summary: | Glucosinolates were resolved with a Kinetex C18 column (2.1 × 50 mm, 2.6 μm) using a flow rate of 0.3 mL min −1 with a mobile phase of 0.1% formic acid in water (A) and 0.1% formic acid in acetonitrile (B). The following gradient was used to elute the analytes: mobile phase B was held at 0% for 1.25 min before increasing to 50% over 2.75 min and to 100% over the next 0.5 min. Solvent B was then held at 100% for 2 min before returning to 0% over 0.5 min for 4 min of column re-equilibration. |
Instrument Name: | Thermo Dionex Ultimate 3000 RS |
Column Name: | Phenomonex Kinetex C18 (50 x 2.1mm,2.6um) |
Column Temperature: | RT |
Flow Rate: | 0.3 ml/min |
Injection Temperature: | 4 |
Solvent A: | 100% water; 0.1% formic acid |
Solvent B: | 100% acetonitrile; 0.1% formic acid |
Chromatography Type: | Reversed phase |