MB Sample ID: SA247845
Local Sample ID: | NHP20 |
Subject ID: | SU002565 |
Subject Type: | Human |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004042 | AN004043 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | HILIC | HILIC |
Chromatography system | Agilent 1290 Infinity | Agilent 1290 Infinity |
Column | Waters ACQUITY UPLC HSS T3 (150 x 2.1mm,1.7um) | Waters ACQUITY UPLC HSS T3 (150 x 2.1mm,1.7um) |
MS Type | ESI | ESI |
MS instrument type | Triple TOF | Triple quadrupole |
MS instrument name | ABI Sciex 6600 TripleTOF | ABI Sciex 6600 TripleTOF |
Ion Mode | POSITIVE | NEGATIVE |
Units | Original peak intensity values | Original peak intensity values |
Chromatography:
Chromatography ID: | CH002990 |
Instrument Name: | Agilent 1290 Infinity |
Column Name: | Waters ACQUITY UPLC HSS T3 (150 x 2.1mm,1.7um) |
Column Temperature: | 25 |
Flow Gradient: | The gradient elution program was as follows: 0 --- 0.5 min, 95% B; 0.5 --- 7 min, B varied linearly from 95% to 65 %; 7 --- 8 min, B varied linearly from 65% to 40%; 8---9 min, B was maintained at 40%; 9---9.1 min, B varied linearly from 40% to 95%; 9.1--12 min, B maintained at 95%. |
Flow Rate: | 0.5 mL/min |
Solvent A: | 100% water; 25 mM ammonium acetate; 25 mM ammonia |
Solvent B: | 100% acetonitrile |
Chromatography Type: | HILIC |