Combined analysis:

Analysis ID AN004405 AN004406
Analysis type MS MS
Chromatography type HILIC HILIC
Chromatography system Agilent 1260 Agilent 1260
Column Waters Atlantis HILIC (150 x 2.1mm,3um) Waters XBridge Amide (100 x 2.1mm,3.5um)
MS Type ESI ESI
MS instrument type QTRAP QTRAP
MS instrument name ABI Sciex 5500 QTrap ABI Sciex 5500 QTrap
Ion Mode POSITIVE POSITIVE
Units Relative abundance Relative abundance

Chromatography:

Chromatography ID:CH003306
Chromatography Summary:HPLC gradient: The HPLC gradient program begins with an initial condition of 5% solvent A and 95% solvent B, with a flow rate of 0.25 mL/min, which is held for 0.5 minutes to establish system equilibration. The gradient then proceeds as follows: at 6 minutes, the mobile phase composition shifts to 60% solvent A and 40% solvent B for 3minutes; at 10 minutes, it changes back to 5% solvent A and 95% solvent B; at 11 minutes, the flow rate increases to 0.4 mL/min while maintaining a composition of 5% solvent A and 95% solvent B for a duration of 12.5 minutes; and at 24.5 minutes, the flow rate decreases back to 0.25 mL/min. The final condition is maintained for 1 minutes to ensure stability before subsequent analyses.
Instrument Name:Agilent 1260
Column Name:Waters Atlantis HILIC (150 x 2.1mm,3um)
Column Temperature:40°C
Flow Gradient:HPLC gradient: The HPLC gradient program begins with an initial condition of 5% solvent A and 95% solvent B, with a flow rate of 0.25 mL/min, which is held for 0.5 minutes to establish system equilibration. The gradient then proceeds as follows: at 6 minutes, the mobile phase composition shifts to 60% solvent A and 40% solvent B for 3minutes; at 10 minutes, it changes back to 5% solvent A and 95% solvent B; at 11 minutes, the flow rate increases to 0.4 mL/min while maintaining a composition of 5% solvent A and 95% solvent B for a duration of 12.5 minutes; and at 24.5 minutes, the flow rate decreases back to 0.25 mL/min.
Flow Rate:0.250 – 0.400 mL/min
Solvent A:0.1% Formic acid in 10 mM Ammonium Formate (pH ~2.5)
Solvent B:0.1% Formic Acid in Acetonitrile
Chromatography Type:HILIC
  
Chromatography ID:CH003307
Chromatography Summary:HPLC gradient: The HPLC gradient program is initialized with an initial mobile phase composition of 15% solvent A and 85% solvent B at a flow rate of 0.25 mL/min. Over the course of 8 minutes, the mobile phase composition undergoes a transition to 65% solvent A and 35% solvent B. Subsequently, at 8 minutes, the composition shifts to 98% solvent A and 2% solvent B, maintained for 1 minute. The mobile phase reverts back to the initial composition of 15% solvent A and 85% solvent B at 10 minutes. At 12.5 minutes, the flow rate is increased to 0.5 mL/min, while maintaining a constant mobile phase composition of 15% solvent A and 85% solvent B for a period of 2.5 minutes. Finally, at 15 minutes, the flow rate is reduced back to 0.25 mL/min. To ensure system stability, the final condition is maintained for 1 minute prior to subsequent analyses.
Instrument Name:Agilent 1260
Column Name:Waters XBridge Amide (100 x 2.1mm,3.5um)
Column Temperature:40°C
Flow Gradient:HPLC gradient: The HPLC gradient program is initialized with an initial mobile phase composition of 15% solvent A and 85% solvent B at a flow rate of 0.25 mL/min. Over the course of 8 minutes, the mobile phase composition undergoes a transition to 65% solvent A and 35% solvent B. Subsequently, at 8 minutes, the composition shifts to 98% solvent A and 2% solvent B, maintained for 1 minute. The mobile phase reverts back to the initial composition of 15% solvent A and 85% solvent B at 10 minutes. At 12.5 minutes, the flow rate is increased to 0.5 mL/min, while maintaining a constant mobile phase composition of 15% solvent A and 85% solvent B for a period of 2.5 minutes.
Flow Rate:0.250 – 0.400 mL/min
Solvent A:95:5 H2O:Acetonitrile (v:v) with 20mM Ammonium Acetate and 20mM Ammonium Hydroxide (pH 9.0)
Solvent B:Acetonitrile
Chromatography Type:HILIC