Return to study ST002736 main page

MB Sample ID: SA288710

Local Sample ID:RJ_17KL1
Subject ID:SU002843
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004438
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Dionex Ultimate 3000
Column SeQuant ZIC- pHILIC (150 x 2.1mm,5um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode UNSPECIFIED
Units peak area

Chromatography:

Chromatography ID:CH003334
Chromatography Summary:HILIC chromatographic separation of metabolites was achieved using a Millipore Sequant ZIC-pHILIC analytical column (5 µm, 2.1 × 150 mm) equipped with a 2.1 × 20 mm guard column (both 5 mm particle size) with a binary solvent system. Solvent A was 20 mM ammonium carbonate, 0.05% ammonium hydroxide; Solvent B was acetonitrile. The column oven and autosampler tray were kept at 40 °C and 4 °C, respectively. The chromatographic gradient was run at a flow rate of 0.200 mL/min as follows: 0–2 min: 80% B; 2-17 min: linear gradient from 80% B to 20% B; 17-17.1 min: linear gradient from 20% B to 80% B; 17.1-22.5 min: hold at 80% B. Samples were randomized and analysed with LC–MS in a blinded manner with an injection volume was 5 µl. Pooled samples were generated from an equal mixture of all individual samples and analysed interspersed, at regular intervals, within the sample sequence as a quality control. Each sample was analysed with three analytical replicates.
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:SeQuant ZIC- pHILIC (150 x 2.1mm,5um)
Column Temperature:40
Flow Gradient:0–2 min: 80% B; 2-17 min: linear gradient from 80% B to 20% B; 17-17.1 min: linear gradient from 20% B to 80% B; 17.1-22.5 min: hold at 80% B
Flow Rate:0.200 mL/min
Solvent A:20 mM ammonium carbonate, 0.05% ammonium hydroxide
Solvent B:acetonitrile
Chromatography Type:HILIC
  logo