MB Sample ID: SA289542

Local Sample ID:Amlodipine 04-02-10453
Subject ID:SU002862
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004472
Analysis type MS
Chromatography type Reversed phase
Chromatography system Bruker Elute
Column Hamilton Intensity Solo 2 C18(100 x 2.1 mm, 1.8 um)
MS Type ESI
MS instrument type QTOF
MS instrument name Bruker timsTOF
Ion Mode POSITIVE
Units AU

Chromatography:

Chromatography ID:CH003357
Chromatography Summary:amples were chromatographically separated by inline reversed-phase chromatography using the Elute HPG 1300 pumps and Elute Autosampler (Bruker, Darmstadt, Germany) with solvent A 0.1% FA in HPLC grade water and solvent B 0.1% FA in ACN. A Hamilton Intensity Solo 2 C18 column (100 mm x 2.1 mm, 1.8µm beads) was maintained at 35C. For metabolomics, 10 µL was injected twice for each sample and eluted using a 30-minute gradient as follows: 1% ACN was held for 2 minutes, ramping to 99% ACN over 15 minutes, held at 99% ACN for 3 minutes before re-equilibrating to 1% ACN for 10 minutes. Flow rates were 250 µL/min for elution and 350 µL/min for re-equilibration.
Instrument Name:Bruker Elute
Column Name:Hamilton Intensity Solo 2 C18(100 x 2.1 mm, 1.8 um)
Column Temperature:35
Flow Gradient:The gradient program was: 0–2 min, 99% A: 1% B; 2–17 min, 99–1% A: 1–99% B; 17–20 min, 99% B: 1% A. The flow rate was fixed at 0.25 mL/min. Subsequently, 20–20.1 min 99% B to 99% A; 20.1–28.5 min, 99% A: 1% B at 0.35 mL/min flow rate; 28.5–30 min; 99% A: 1% B at 0.25 mL/min.
Flow Rate:250 uL/min
Solvent A:Water (0.1% Formic Acid)
Solvent B:ACN (0.1% Formic Acid)
Chromatography Type:Reversed phase