Return to study ST002773 main page

MB Sample ID: SA294727

Local Sample ID:SWHS0899
Subject ID:SU002880
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004513 AN004514
Analysis type MS MS
Chromatography type HILIC Reversed phase
Chromatography system Thermo Dionex Ultimate 3000 RS Thermo Dionex Ultimate 3000 RS
Column Higgins Analytical TARGA C18 (50 x 2.1mm,5um) Waters XBridge BEH Amide (50 x 2.1mm,2.5um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE NEGATIVE
Units Peak Intensity Peak Intensity

Chromatography:

Chromatography ID:CH003390
Chromatography Summary:The HILIC column is operated parallel to reverse phase column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10-port and 6-port switching valves. During operation of HILIC separation method, the MS is operated in positive ion mode and 10 μL of sample is injected onto the HILIC column while the reverse phase column is flushing with wash solution. Flow rate is maintained at 0.35 mL/min until 1.5 min, increased to 0.4 mL/min at 4 min and held for 1 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 2% formic acid (v/v) in LC-MS grade water. Initial mobile phase conditions are 22.5% A, 75% B, 2.5% C hold for 1.5 min, with linear gradient to 77.5% A, 20% B, 2.5% C at 4 min, hold for 1 min, resulting in a total analytical run time of 5 min. During the flushing phase (reverse phase analytical separation), the HILIC column is equilibrated with a wash solution of 77.5% A, 20% B, 2.5% C.
Methods ID:EmoryUniversity_HRM_DC5min_082016_01.pdf
Instrument Name:Thermo Dionex Ultimate 3000 RS
Column Name:Higgins Analytical TARGA C18 (50 x 2.1mm,5um)
Column Temperature:40
Flow Gradient:Initial mobile phase conditions are 22.5% A, 75% B, 2.5% C hold for 1.5 min, with linear gradient to 77.5% A, 20% B, 2.5% C at 4 min, hold for 1 min, resulting in a total analytical run time of 5 min. During the flushing phase (reverse phase analytical separation), the HILIC column is equilibrated with a wash solution of 77.5% A, 20% B, 2.5% C.
Flow Rate:0.35-0.4 mL/min
Solvent A:100% water
Solvent B:100% acetonitrile
Chromatography Type:HILIC
Solvent C:100% water; 2% formic acid
  
Chromatography ID:CH003391
Chromatography Summary:The C18 column is operated parallel to the HILIC column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10-port and 6-port switching valves. During operation of the C18 method, the MS is operated in negative ion mode and 10 μL of sample is injected onto the C18 column while the HILIC column is flushing with wash solution. Flow rate is maintained at 0.4 mL/min until 1.5 min, increased to 0.5 mL/min at 2 min and held for 3 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 10mM ammonium acetate in LC-MS grade water. Initial mobile phase conditions are 60% A, 35% B, 5% C hold for 0.5 min, with linear gradient to 0% A, 95% B, 5% C at 1.5 min, hold for 3.5 min, resulting in a total analytical run time of 5 min. During the flushing phase (HILIC analytical separation), the C18 column is equilibrated with a wash solution of 0% A, 95% B, 5% C until 2.5 min, followed by an equilibration solution of 60% A, 35% B, 5% C for 2.5 min.
Methods ID:EmoryUniversity_HRM_DC5min_082016_01.pdf
Instrument Name:Thermo Dionex Ultimate 3000 RS
Column Name:Waters XBridge BEH Amide (50 x 2.1mm,2.5um)
Column Temperature:40
Flow Gradient:Initial mobile phase conditions are 60% A, 35% B, 5% C hold for 0.5 min, with linear gradient to 0% A, 95% B, 5% C at 1.5 min, hold for 3.5 min, resulting in a total analytical run time of 5 min. During the flushing phase (HILIC analytical separation), the C18 column is equilibrated with a wash solution of 0% A, 95% B, 5% C until 2.5 min, followed by an equilibration solution of 60% A, 35% B, 5% C for 2.5 min.
Flow Rate:0.4-0.5 mL/min
Solvent A:100% water
Solvent B:100% acetonitrile
Chromatography Type:Reversed phase
Solvent C:100% water; 10 mM ammonium acetate