MB Sample ID: SA306584
Local Sample ID: | M00179_20220322_Neg_1_MSA0041 |
Subject ID: | SU002941 |
Subject Type: | Bacteria |
Subject Species: | Bacteroides thetaiotaomicron |
Taxonomy ID: | 8188 |
Subject Comments: | Fecal derived communities and isolates, supernatant was assayed |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004625 | AN004626 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | HILIC | HILIC |
Chromatography system | Thermo Vanquish | Thermo Vanquish |
Column | Waters Acquity BEH Amide (150 x 2.1mm, 1.7um) | Waters Acquity BEH Amide (150 x 2.1mm, 1.7um) |
MS Type | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive HF hybrid Orbitrap | Thermo Q Exactive HF hybrid Orbitrap |
Ion Mode | POSITIVE | NEGATIVE |
Units | counts, height | counts, height |
Chromatography:
Chromatography ID: | CH003481 |
Chromatography Summary: | Bacterial supernatant were analyzed via hydrophilic interaction liquid chromatography (HILIC) coupled to a Thermo Q-Exactive HF high resolution mass spectrometer. Each sample was analyzed in both positive and negative ionization modes (ESI+, ESI-) via subsequent injections. Full MS-ddMS2 data was collected, an inclusion list was used to prioritize MS2 selection of metabolites from our in-house ‘local’ library, when additional scan bandwidth was available MS2 was collected in a data-dependent manner. Mass range was 60-900 mz, resolution was 60k (MS1) and 15k (MS2), centroid data was collected, loop count was 4, isolation window was 1.5 Da. Metabolomics data was processed using MS-DIAL v4.60 (https://www.nature.com/articles/s41587-020-0531-2) and queried against a combination of our in-house MS2 library (https://www.nature.com/articles/s41586-021-03707-9) and MassBank of North America, the largest freely available spectral repository (https://doi.org/10.1002/mas.21535). Annotations were scored using guidelines from the metabolomics standards initiative (https://www.nature.com/articles/nbt0807-846b). Features were excluded from analysis if peak height was not at least 5-fold greater in one or more samples compared to the procedural blank average. |
Instrument Name: | Thermo Vanquish |
Column Name: | Waters Acquity BEH Amide (150 x 2.1mm, 1.7um) |
Column Temperature: | 45C |
Flow Gradient: | Gradient elution was performed from 100% (B) at 0–2 min to 70% (B) at 7.7 min, 40% (B) at 9.5 min, 30% (B) at 10.25 min, 100% (B) at 12.75 min, isocratic until 16.75 min with a column flow of |
Flow Rate: | 0.4 mL/min. |
Solvent A: | Water + 10mM ammonium formate + 0.125% formic acid |
Solvent B: | 95% acetonitrile + 10mM ammonium formate + 0.125% formic acid |
Chromatography Type: | HILIC |