MB Sample ID: SA306598

Local Sample ID:M00179_20220322_Pos_1_MSA0072
Subject ID:SU002941
Subject Type:Bacteria
Subject Species:Bacteroides thetaiotaomicron
Taxonomy ID:8188
Subject Comments:Fecal derived communities and isolates, supernatant was assayed

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004625 AN004626
Analysis type MS MS
Chromatography type HILIC HILIC
Chromatography system Thermo Vanquish Thermo Vanquish
Column Waters Acquity BEH Amide (150 x 2.1mm, 1.7um) Waters Acquity BEH Amide (150 x 2.1mm, 1.7um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE NEGATIVE
Units counts, height counts, height

Chromatography:

Chromatography ID:CH003481
Chromatography Summary:Bacterial supernatant were analyzed via hydrophilic interaction liquid chromatography (HILIC) coupled to a Thermo Q-Exactive HF high resolution mass spectrometer. Each sample was analyzed in both positive and negative ionization modes (ESI+, ESI-) via subsequent injections. Full MS-ddMS2 data was collected, an inclusion list was used to prioritize MS2 selection of metabolites from our in-house ‘local’ library, when additional scan bandwidth was available MS2 was collected in a data-dependent manner. Mass range was 60-900 mz, resolution was 60k (MS1) and 15k (MS2), centroid data was collected, loop count was 4, isolation window was 1.5 Da. Metabolomics data was processed using MS-DIAL v4.60 (https://www.nature.com/articles/s41587-020-0531-2) and queried against a combination of our in-house MS2 library (https://www.nature.com/articles/s41586-021-03707-9) and MassBank of North America, the largest freely available spectral repository (https://doi.org/10.1002/mas.21535). Annotations were scored using guidelines from the metabolomics standards initiative (https://www.nature.com/articles/nbt0807-846b). Features were excluded from analysis if peak height was not at least 5-fold greater in one or more samples compared to the procedural blank average.
Instrument Name:Thermo Vanquish
Column Name:Waters Acquity BEH Amide (150 x 2.1mm, 1.7um)
Column Temperature:45C
Flow Gradient:Gradient elution was performed from 100% (B) at 0–2 min to 70% (B) at 7.7 min, 40% (B) at 9.5 min, 30% (B) at 10.25 min, 100% (B) at 12.75 min, isocratic until 16.75 min with a column flow of
Flow Rate:0.4 mL/min.
Solvent A:Water + 10mM ammonium formate + 0.125% formic acid
Solvent B:95% acetonitrile + 10mM ammonium formate + 0.125% formic acid
Chromatography Type:HILIC