MB Sample ID: SA313392

Local Sample ID:serum-AK-E18-C
Subject ID:SU002983
Subject Type:Mammal
Subject Species:Mus musculus
Taxonomy ID:10090

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004705 AN004706
Analysis type MS MS
Chromatography type HILIC HILIC
Chromatography system Thermo Vanquish Thermo Vanquish
Column SeQuant ZIC-HILIC (150 x 2.1mm,5um) SeQuant ZIC-HILIC (150 x 2.1mm,5um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units Peak Area Peak Area

Chromatography:

Chromatography ID:CH003543
Chromatography Summary:Dried metabolites were reconstituted in 100 µL of a 50% acetonitrile (ACN) 50% dH20 solution. Samples were vortexed and spun down for 10 min at 17,000g. 70 µL of the supernatant was then transferred to HPLC glass vials. 10 µL of these metabolite solutions were injected per analysis. Samples were run on a Vanquish (Thermo Scientific) UHPLC system with mobile phase A (20mM ammonium carbonate, pH 9.7) and mobile phase B (100% ACN) at a flow rate of 150 µL/min on a SeQuant ZIC-pHILIC Polymeric column (2.1 × 150 mm 5 μm, EMD Millipore) at 35°C. Separation was achieved with a linear gradient from 20% A to 80% A in 20 min followed by a linear gradient from 80% A to 20% A from 20 min to 20.5 min. 20% A was then held from 20.5 min to 28 min.
Instrument Name:Thermo Vanquish
Column Name:SeQuant ZIC-HILIC (150 x 2.1mm,5um)
Column Temperature:35°C
Flow Gradient:Separation was achieved with a linear gradient from 20% A to 80% A in 20 min followed by a linear gradient from 80% A to 20% A from 20 min to 20.5 min. 20% A was then held from 20.5 min to 28 min.
Flow Rate:150 µL/min
Solvent A:100% water; 20 mM ammonium carbonate, pH 9.7
Solvent B:100% acetonitrile
Chromatography Type:HILIC