MB Sample ID: SA315649

Local Sample ID:Alm 30.3
Subject ID:SU003015
Subject Type:Food item

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004761
Analysis type MS
Chromatography type GC
Chromatography system Leco Pegasus BT
Column Zebron ZB-FAME (30m × 0.25mm, 0.2um)
MS Type EI
MS instrument type GC-TOF
MS instrument name Agilent 7890B
Ion Mode POSITIVE
Units uMol

Chromatography:

Chromatography ID:CH003593
Chromatography Summary:The upper phase containing FAMEs was taken for analysis using a GC coupled to a time of flight mass spectrometry (GC-TOFMS, Pegasus BT, Leco Corp., St.Joseph, MI) on a ZB-FAME (30 m ×0.25 mm ×0.2 μm, Zebron) in EI mode. One microliter of sample was injected with a split ratio of 60 to 1. Helium was used as a carrier gas. The total flow was 1 mL min−1. The GC oven temperature was programmed from 40 °C (2 min hold) to 150 °C at 20 °C min−1 (1 min hold), to 165 °C at 1 °C min−1 (1 min hold), to 180 °C at 2 °C min−1, to 200 °C at 5 °C min−1, and to 260 °C at 40 °C min−1 (8 min hold). The transfer line and ion source temperatures were kept at 240 °C and 220 °C, respectively. The TOFMS scanned from m/z 45–650 with a data acquisition rate of 20 spectra s−1. MS data was acquired and processed using ChromaTOF software (version 5.32, Leco Corp.).
Instrument Name:Leco Pegasus BT
Column Name:Zebron ZB-FAME (30m × 0.25mm, 0.2um)
Column Temperature:40
Flow Gradient:na
Flow Rate:1
Solvent A:na
Solvent B:na
Chromatography Type:GC