Return to study ST003114 main page

MB Sample ID: SA337394

Local Sample ID:S12_mitos_control_4
Subject ID:SU003230
Subject Type:Cultured cells
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN005102 AN005103 AN005104 AN005105
Analysis type MS MS MS MS
Chromatography type None (Direct infusion) Normal phase Reversed phase Reversed phase
Chromatography system Advion TriVersa NanoMate Agilent 1260 Shimadzu Nexera X2 Shimadzu Nexera X2
Column None Macherey-Nagel Nucleosil NH2 (50×2 mm, 3 um, 120 Å) Waters Acquity BEH Shield RP18 (100×2.1 mm, 1.7 um) Phenomenex Core-Shell Kinetex Biphenyl (100×3.0 mm, 2.6 um, 100 Å)
MS Type ESI ESI ESI ESI
MS instrument type QTRAP QTRAP QTRAP QTRAP
MS instrument name SCIEX QTRAP 6500 SCIEX QTRAP 6500 SCIEX QTRAP 6500 SCIEX QTRAP 6500
Ion Mode POSITIVE POSITIVE POSITIVE NEGATIVE
Units counts per second (cps) counts per second (cps) counts per second (cps) counts per second (cps)

Chromatography:

Chromatography ID:CH003858
Chromatography Summary:Lipid extract infusion and ionization was conducted using Nano-ESI chips with the TriVersa NanoMate operated by the ChipSoft Software (Advion) under the following settings: sample infusion volume: 14 μl, volume of air to aspirate after sample: 1 μl, air gap before chip: enabled, aspiration delay: 0 s, prepiercing: with mandrel, spray sensing: enabled, cooling temperature: 14°C, gas pressure: 0.5 psi, ionization voltage: 1.4 kV, and vent headspace: enabled. Prewetting was done once.
Instrument Name:Advion TriVersa NanoMate
Column Name:None
Column Temperature:N/A
Flow Gradient:N/A
Flow Rate:N/A
Solvent A:None
Solvent B:None
Chromatography Type:None (Direct infusion)
  
Chromatography ID:CH003859
Chromatography Summary:Note: Macherey-Nagel does not offer this column as standard, but manufactures it on request.
Instrument Name:Agilent 1260
Column Name:Macherey-Nagel Nucleosil NH2 (50×2 mm, 3 um, 120 Å)
Column Temperature:20
Flow Gradient:0 min: 0% B, 0.5 min: 0% B, 0.7 min: 10% B, 1.2 min: 10% B, 1.6 min: 18% B, 2.2 min: 18% B, 2.6 min: 100% B, 4.5 min: 100% B, 4.9 min: 0% B, 6.5 min: 0% B
Flow Rate:0.75 ml/min
Solvent A:97% acetonitrile/2% methanol/1% acetic acid; 5 mM ammonium acetate
Solvent B:99% methanol/1% acetic acid; 5 mM ammonium acetate
Chromatography Type:Normal phase
  
Chromatography ID:CH003860
Instrument Name:Shimadzu Nexera X2
Column Name:Waters Acquity BEH Shield RP18 (100×2.1 mm, 1.7 um)
Column Temperature:50
Flow Gradient:0 min: 30% B, 0.5 min: 30% B, 4.5 min: 68% B, 20.5 min: 75% B, 21 min: 97% B, 24 min: 97% B, 24.5 min: 30% B, 28 min: 30% B
Flow Rate:0.4 ml/min
Solvent A:60% acetonitrile/40% water; 10 mM ammonium formate
Solvent B:90% isopropanol/10% acetonitrile; 10 mM ammonium formate
Chromatography Type:Reversed phase
  
Chromatography ID:CH003861
Instrument Name:Shimadzu Nexera X2
Column Name:Phenomenex Core-Shell Kinetex Biphenyl (100×3.0 mm, 2.6 um, 100 Å)
Column Temperature:40
Flow Gradient:0 min: 55% B, 4 min: 95% B, 7 min: 95% B, 7.1 min: 55% B, 10 min: 55% B
Flow Rate:0.7 ml/min
Solvent A:100% water; 0.012 % acetic acid; 5 mM ammonium acetate
Solvent B:80% acetonitrile/20% isopropanol
Chromatography Type:Reversed phase
  logo