This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001773. The data can be accessed directly via it's Project DOI: 10.21228/M8J420 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST002907
Study TitleFolate levels in K562 cells following short-term folate depletion
Study SummaryCulture of K562 cells for 48hr in RPMI media containing 2000 nM, 100 nM, 50 nM, 20 nM, or 0 nM folic acid followed by folate metabolomics.
Institute
Boston Children's Hospital, Harvard Medical School
Departmentpathology
LaboratoryKanarek Lab
Last NameKanarek
First NameNaama
AddressEnders 1116.2, 300 Longwood Ave, Boston, MA 02115
Emailnaama.kanarek@childrens.harvard.edu
Phone6173557433
Submit Date2023-10-02
Num Groups6
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2024-02-13
Release Version1
Naama Kanarek Naama Kanarek
https://dx.doi.org/10.21228/M8J420
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004770
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Vanquish
Column Phenomenex Kinetex C18 (150 x 3mm,2.6um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode UNSPECIFIED
Units peak area

Chromatography:

Chromatography ID:CH003602
Chromatography Summary:5 μl was injected onto a 2.6 μm, 150 x 3 mm C18 column (Phenomenex, 00F-4462-Y0) equipped with a 3.0 mm safe-guard column (Phenomenex, AJ0-8775). The column oven and autosampler tray were held at 30 °C and 4 °C, respectively. The following conditions were used to achieve chromatographic separation: buffer A was 0.1% formic acid; buffer B was acetonitrile with 0.1% formic acid. The chromatographic gradient was run at a flow rate of 0.250 ml min−1 as follows: 0–5 min: gradient was held at 5% B; 5–10 min: linear gradient of 5% to 36% B; 10.1–14.0 min: linear gradient from 36–95% B; 14.1–18.0 min: gradient was returned to 5% B.
Instrument Name:Thermo Vanquish
Column Name:Phenomenex Kinetex C18 (150 x 3mm,2.6um)
Column Temperature:30
Flow Gradient:0.250 ml min−1 as follows: 0–5 min: gradient was held at 5% B; 5–10 min: linear gradient of 5% to 36% B; 10.1–14.0 min: linear gradient from 36–95% B; 14.1–18.0 min: gradient was returned to 5% B.
Flow Rate:250 uL/min
Solvent A:100% water; 0.1% formic acid
Solvent B:100% acetonitrile; 0.1% formic acid
Chromatography Type:Reversed phase