Summary of Study ST000626
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000456. The data can be accessed directly via it's Project DOI: 10.21228/M83W49 This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST000626 |
Study Title | Non targeted metabolomic analysis of mouse and human bronchoalveolar lavage fluid, Lipid(-) experiment (part II) |
Study Type | Comprehensive profile |
Study Summary | Mouse BALF was collected from C57BL/6 mice (n = 10). Five mice were exposed to ambient air for one day (air control) and five mice were exposed to CS for nine months (smoking). Human BALF was collected from subjects from the COPDGene cohort (n = 5). COPD diagnosis was based on the ratio of forced expiratory volume in 1 second to forced vital capacity (FEV1/FVC). Subjects were 45-70 years old, BMI 27-45, weight 76-125kg, and were categorized as follows: 1 male former smoker without COPD (FEV1/FVC = 0.82), 2 male current smokers without COPD (FEV1/FVC = 0.91 and 0.79), 1 female current smoker with moderate COPD (FEV1/FVC = 0.51), and 1 male current smoker with moderate COPD (FEV1/FVC = 0.64) |
Institute | University of Colorado Denver |
Department | Anschutz Medical Campus, Pharmaceutical Sciences |
Laboratory | Metabolomics and Mass Spectrometry |
Last Name | Reisdorph |
First Name | Nichole |
Address | 12850 East Montview Blvd, Denver, CO 80204 |
nichole.reisdorph@ucdenver.edu | |
Phone | - |
Submit Date | 2017-04-21 |
Num Groups | 2 |
Total Subjects | 36 |
Raw Data Available | Yes |
Raw Data File Type(s) | d |
Analysis Type Detail | LC-MS |
Release Date | 2018-12-11 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN000958 |
---|---|
Analysis type | MS |
Chromatography type | Reversed phase |
Chromatography system | Agilent 1290 |
Column | Agilent Zorbax RRHD SB-C18 (100 x 2.1mm,1.8um) |
MS Type | ESI |
MS instrument type | QTOF |
MS instrument name | Agilent 6510 QTOF |
Ion Mode | NEGATIVE |
Units | peak intensity |
Chromatography:
Chromatography ID: | CH000683 |
Chromatography Summary: | The BALF samples were separated using C18 reverse phase chromatography. The separations used an Agilent Zorbax Rapid Resolution HD (RRHD) SB-C18, 1.8 micron (2.1 x 100 mm) analytical column with an Agilent Zorbax SB-C18, 1.8 micron (2.1 x 5 mm) guard column and used an Agilent 1290 series high performance liquid chromatography (HPLC) pump. Injection volumes were 4µL for mouse BALF and 15µL for human BALF. Flow rate was 0.7 mL/min using the following mobile phases: mobile phase A was water with 0.1% formic acid, and mobile phase B was 60:36:4 isopropyl alcohol:acetonitrile:water with 0.1% formic acid. Gradients were 0-0.5 minutes 30-70% B, 0.5-7.42 minutes 70-100% B, 7.42-9.9 minutes 100% B, 9.9-10.0 minutes 100-30% B, 10-14.6 minutes 30% B for the positive mode injections. For negative mode the following eluant profile was used: 0-1 minutes 30-70% B, 1-7.92 minutes 70-100% B, 7.92-10.4 minutes 100% B, 10.4-10.5 minutes 100-30% B, 10.5-15.1 minutes 30% B. |
Instrument Name: | Agilent 1290 |
Column Name: | Agilent Zorbax RRHD SB-C18 (100 x 2.1mm,1.8um) |
Flow Gradient: | 0-1 minutes 30-70% B, 1-7.92 minutes 70-100% B, 7.92-10.4 minutes 100% B, 10.4-10.5 minutes 100-30% B, 10.5-15.1 minutes 30% B. |
Flow Rate: | 0.7 mL/min |
Solvent A: | 100% water; 0.1% formic acid |
Solvent B: | 60% isopropanol/36% acetonitrile/4% water; 0.1% formic acid |
Chromatography Type: | Reversed phase |