Summary of Study ST002790

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001021. The data can be accessed directly via it's Project DOI: 10.21228/M8QM5H This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Show all samples  |  Perform analysis on untargeted data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST002790
Study TitleCommunity metabolomes reflect taxon-specific fingerprints of phytoplankton and heterotrophic bacteria in the ocean (expanded)
Study TypeMetabolomic survey of 46 phytoplankton and heterotrophic bacteria species
Study SummaryThis study is an updated version of the prior study in this project, ST001514. With this update we introduce additional marine species, in particular the heterotrophic bacteria, and include our expanded library of authentic standards (available at https://github.com/IngallsLabUW/Ingalls_Standards/blob/430992e10c0464a817c69d22e3905d73f2ea196f/Ingalls_Lab_Standards.csv) for improved targeted coverage.
Institute
University of Washington
Last NameKumler
First NameWilliam
Address1501 NE Boat St, Seattle, WA, 98105, USA
Emailwkumler@uw.edu
Phone9095551234
Submit Date2023-07-20
Raw Data AvailableYes
Raw Data File Type(s)mzML
Analysis Type DetailLC-MS
Release Date2023-08-07
Release Version1
William Kumler William Kumler
https://dx.doi.org/10.21228/M8QM5H
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004540
Analysis type MS
Chromatography type HILIC
Chromatography system Waters Acquity I-Class
Column SeQuant ZIC-pHILIC (150 x 2.1mm,5um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE
Units normalized peak area per liter seawater filtered

Chromatography:

Chromatography ID:CH003410
Chromatography Summary:A SeQuant ZIC-pHILIC column (5 um particle size, 2.1 mm x 150 mm, from Millipore) was used with 10 mM ammonium carbonate in 85:15 acetonitrile to water (Solvent A) and 10 mM ammonium carbonate in 85:15 water to acetonitrile (Solvent B) at a flow rate of 0.15 mL/min. This column was compared with a Waters UPLC BEH amide and a Millipore cHILIC column; the pHILIC showed superior reproducibility and peak shapes. The column was held at 100% A for 2 minutes, ramped to 64% B over 18 minutes, ramped to 100% B over 1 minute, held at 100% B for 5 minutes, and equilibrated at 100% A for 25 minutes (50 minutes total). The column was maintained at 30 C. The injection volume was 2 µL for samples and standard mixes. When starting a batch, the column was equilibrated at the starting conditions for at least 30 minutes. To improve the performance of the HILIC column, we maintained the same injection volume, kept the instrument running water blanks between samples as necessary, and injected standards in a representative matrix in addition to standards in water. After each batch, the column was flushed with 10 mM ammonium carbonate in 85:15 water to acetonitrile for 20 to 30 minutes.
Instrument Name:Waters Acquity I-Class
Column Name:SeQuant ZIC-pHILIC (150 x 2.1mm,5um)
Column Temperature:30
Flow Gradient:100% A for 2 minutes, ramped to 64% B over 18 minutes, ramped to 100% B over 1 minute, held at 100% B for 5 minutes, and equilibrated at 100% A for 25 minutes (50 minutes total)
Flow Rate:0.15mL/min
Solvent A:85% acetonitrile/15% water; 10 mM ammonium carbonate
Solvent B:15% acetonitrile/85% water; 10 mM ammonium carbonate
Chromatography Type:HILIC