Return to study ST001811 main page

MB Sample ID: SA168262

Local Sample ID:WTp2GLN-3-neg-PRM
Subject ID:SU001888
Subject Type:Plant
Subject Species:Arabidopsis thaliana
Taxonomy ID:3702
Genotype Strain:Col-0
Age Or Age Range:10 day old seedlings
Gender:Not applicable

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002935 AN002936
Analysis type MS MS
Chromatography type HILIC HILIC
Chromatography system Agilent 1290 Infinity II Agilent 1290 Infinity II
Column SeQuant ZIC-HILIC (100 x 2.1mm,3.5um) SeQuant ZIC-HILIC (100 x 2.1mm,3.5um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units umol g FW-1 umol g FW-1

MS:

MS ID:MS002726
Analysis ID:AN002935
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The following heated electrospray ionization (HESI) conditions were optimized for the analysis of amino and organic acids: spray voltage, 3.9 kV (ESI+), 3.5 kV (ESI-); capillary temperature, 250 °C; probe heater temperature, 450 °C; sheath gas, 30 arbitrary units; auxiliary gas, 8 arbitrary units; and S-Lens RF level, 60%. Injections of 5 μl were used with a flow rate of 0.3 mL min-1. Compounds were detected and monitored using targeted MS/MS, spectra were collected at 17,500 resolution, AGC target 1e6, maximum IT 65 ms, isolation window of 1 m/z, normalized collision energy of 30, intensity threshold of 1.6e5 and 10s dynamic exclusion.
Ion Mode:POSITIVE
Analysis Protocol File:Targeted_Metabolite_Analysis.pdf
  
MS ID:MS002727
Analysis ID:AN002936
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The following heated electrospray ionization (HESI) conditions were optimized for the analysis of amino and organic acids: spray voltage, 3.9 kV (ESI+), 3.5 kV (ESI-); capillary temperature, 250 °C; probe heater temperature, 450 °C; sheath gas, 30 arbitrary units; auxiliary gas, 8 arbitrary units; and S-Lens RF level, 60%. Injections of 5 μl were used with a flow rate of 0.3 mL min-1. Compounds were detected and monitored using targeted MS/MS, spectra were collected at 17,500 resolution, AGC target 1e6, maximum IT 65 ms, isolation window of 1 m/z, normalized collision energy of 30, intensity threshold of 1.6e5 and 10s dynamic exclusion.
Ion Mode:NEGATIVE
Analysis Protocol File:Targeted_Metabolite_Analysis