Combined analysis:

Analysis ID AN003382
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Vanquish
Column Waters Acquity UPLC BEH amide,(150 x 2.1mm,1.7um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE
Units peak height

MS:

MS ID:MS003149
Analysis ID:AN003382
Instrument Name:Thermo Q Exactive HF hybrid Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:A Thermo Q-Exactive HF Orbitrap MS instrument was operated in positive and negative ESI mdoes respectively with the following parameters: mass range 60−900 m/z; spray voltage 3.6kV (ESI+) and −3kV (ESI−), sheath gas (nitrogen) flow rate 60 units; auxiliary gas (nitrogen) flow rate 25 units, capillary temperature 320 ◦C, full scan MS1 mass resolving power 60,000, data-dependent MSMS (dd-MSMS) 4 scans per cycle, normalized collision energy at 20%, 30%, and 40%, dd-MSMS mass resolving power 15,000. Thermo Xcalibur 4.0.27.19 was used for data acquisition and analysis. Instruments was tuned and calibrated by manufacturer’s recommendations. MSDIAL was used for data processing. Conjugated bile acid peak height was converted to relative concentration in ng/mL by comparing to targeted bile acid dataset ST002073.
Ion Mode:POSITIVE