MB Sample ID: SA207487

Local Sample ID:pooledQC 03
Subject ID:SU002248
Subject Type:Mammal
Subject Species:Mus musculus
Taxonomy ID:10090

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003542 AN003543
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Agilent 6550 Agilent 6550
Column Waters Acquity CSH C18 (100 x 2.1mm,1.7um) Waters Acquity CSH C18 (100 x 2.1mm,1.7um)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name Agilent 6540 QTOF Agilent 6545 QTOF
Ion Mode POSITIVE NEGATIVE
Units pmol per sample pmol per sample

MS:

MS ID:MS003300
Analysis ID:AN003542
Instrument Name:Agilent 6540 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:For positive mode, the source gas temperature was set to 225 °C, with a drying gas flow of 11 L/minute, nebulizer pressure of 40 psig, sheath gas temp of 350 °C and sheath gas flow of 11 L/minute. VCap voltage is set at 3500 V, nozzle voltage 500V, fragmentor at 110 V, skimmer at 85 V and octopole RF peak at 750 V. For data processing, Agilent MassHunter (MH) Workstation and software packages MH Qualitiative and MH Quantitative were used. The pooled QC (n=8) and process blank (n=4) were injected throughout the sample queue to ensure the reliability of acquired lipidomics data. For lipid annotation, accurate mass and MS/MS matching was used with the Agilent Lipid Annotator library. Results from the positive and negative ionization modes from Lipid Annotator were merged based on the class of lipid identified. Data exported from MH Quantitative was evaluated using Excel where initial lipid targets are parsed based on the following criteria. Only lipids with relative standard deviations (RSD) less than 30% in QC samples are used for data analysis. Additionally, only lipids with background AUC counts in process blanks that are less than 30% of QC are used for data analysis. The parsed excel data tables are normalized based on the ratio to class-specific internal standards, then to sum prior to statistical analysis.
Ion Mode:POSITIVE
  
MS ID:MS003301
Analysis ID:AN003543
Instrument Name:Agilent 6545 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:For negative mode, the source gas temperature was set to 300 °C, with a drying gas flow of 11 L/minute, a nebulizer pressure of 30 psig, sheath gas temp of 350 °C and sheath gas flow 11 L/minute. VCap voltage was set at 3500 V, nozzle voltage 75 V, fragmentor at 175 V, skimmer at 75 V and octopole RF peak at 750 V. For data processing, Agilent MassHunter (MH) Workstation and software packages MH Qualitiative and MH Quantitative were used. The pooled QC (n=8) and process blank (n=4) were injected throughout the sample queue to ensure the reliability of acquired lipidomics data. For lipid annotation, accurate mass and MS/MS matching was used with the Agilent Lipid Annotator library. Results from the positive and negative ionization modes from Lipid Annotator were merged based on the class of lipid identified. Data exported from MH Quantitative was evaluated using Excel where initial lipid targets are parsed based on the following criteria. Only lipids with relative standard deviations (RSD) less than 30% in QC samples are used for data analysis. Additionally, only lipids with background AUC counts in process blanks that are less than 30% of QC are used for data analysis. The parsed excel data tables are normalized based on the ratio to class-specific internal standards, then to sum prior to statistical analysis.
Ion Mode:NEGATIVE