MB Sample ID: SA265620

Local Sample ID:T2D_GUOU+
Subject ID:SU002783
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004353 AN004354
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Thermo Vanquish Thermo Vanquish
Column Phenomenex Kinetex C18 (100 x 2.1mm,1.7um) Phenomenex Kinetex C18 (100 x 2.1mm,1.7um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units AU AU

MS:

MS ID:MS004100
Analysis ID:AN004353
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The Q Exactive mass spectrometer (ThermoFisher) was operated independently in positive or negative ion mode, scanning in Full MS mode (2 microscans) from 60 to 900 m/z at 70,000 resolution, with 4 kV spray voltage, 45 sheath gas, 15 auxiliary gas. Calibration was performed prior to analysis using the PierceTM Positive and Negative Ion Calibration Solutions (ThermoFisher). Acquired data was then converted from raw to mzXML file format using Mass Matrix (Cleveland, OH). Samples were analyzed in randomized order with a technical mixture injected periodically through analysis to qualify instrument performance. Metabolite assignments, isotopologue distributions, and correction for expected natural abundances of deuterium, 13C, and 15N isotopes were performed using MAVEN (Princeton, NJ)
Ion Mode:POSITIVE
  
MS ID:MS004101
Analysis ID:AN004354
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The Q Exactive mass spectrometer (ThermoFisher) was operated independently in positive or negative ion mode, scanning in Full MS mode (2 microscans) from 60 to 900 m/z at 70,000 resolution, with 4 kV spray voltage, 45 sheath gas, 15 auxiliary gas. Calibration was performed prior to analysis using the PierceTM Positive and Negative Ion Calibration Solutions (ThermoFisher). Acquired data was then converted from raw to mzXML file format using Mass Matrix (Cleveland, OH). Samples were analyzed in randomized order with a technical mixture injected periodically through analysis to qualify instrument performance. Metabolite assignments, isotopologue distributions, and correction for expected natural abundances of deuterium, 13C, and 15N isotopes were performed using MAVEN (Princeton, NJ)
Ion Mode:NEGATIVE