Return to study ST002935 main page

MB Sample ID: SA318402

Local Sample ID:Les-Jas-20190812-13
Subject ID:SU003048
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004814
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Dionex Ultimate 3000
Column Water's Xbridge amide (100 x 3mm, 3.5 um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Plus Orbitrap
Ion Mode UNSPECIFIED
Units Peak Area

MS:

MS ID:MS004560
Analysis ID:AN004814
Instrument Name:Thermo Q Exactive Plus Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The capillary of the electrospray ionization source was set to 275°C, with sheath gas at 45 arbitrary units, auxiliary gas at 5 arbitrary units, and the spray voltage at 4.0 kV. In positive/negative polarity switching mode, a mass/charge ratio (m/z) scan range from 70 to 850 was chosen and MS1 data were collected at a resolution of 70,000. The automatic gain control target was set at 1 × 106, and the maximum injection time was 200 ms. The top five precursor ions were subsequently fragmented, in a data-dependent manner, using the higher-energy collisional dissociation cell set to 30% normalized collision energy in MS2 at a resolution power of 17,500.
Ion Mode:UNSPECIFIED
  logo