Return to study ST000048 main page

MB Sample ID: SA002113

Local Sample ID:A04-0
Subject ID:SU000066
Subject Type:Animal
Subject Species:Mirounga angustirostris
Taxonomy ID:9716
Species Group:Mammal

Select appropriate tab below to view additional metadata details:


Subject:

Subject ID:SU000066
Subject Type:Animal
Subject Species:Mirounga angustirostris
Taxonomy ID:9716
Species Group:Mammal

Factors:

Local Sample IDMB Sample IDFactor Level IDLevel ValueFactor Name
A04-0SA002113FL000647No GLP1/Late GTTTreatment
A04-0SA002113FL0006470Minutes

Collection:

Collection ID:CO000049
Collection Summary:-
Sample Type:Blood

Treatment:

Treatment ID:TR000067

Sample Preparation:

Sampleprep ID:SP000062
Sampleprep Summary:-
Sampleprep Protocol Filename:SOP_Extraction_of_Blood_Samples.pdf
SOP_Derivatization_of_Samples-Standards_2.pdf
Processing Method:Homogenization, protein precipitation via extraction, Solvent Removal w/ Speed Vac
Processing Storage Conditions:On Ice
Extraction Method:3:3:2 Acetonitrile/Isopropanol/Water
Extract Concentration Dilution:30uL of plasma/ml
Extract Cleanup:1:1 Acetonitrile/Water
Extract Storage:-20C Freezer
Sample Derivatization:10ul of 40mg/ml MeOX in pyridine, 91ul of MSTFA +FAMEs mixture
Sample Spiking:FAMEs

Combined analysis:

Analysis ID AN000084
Analysis type MS
Chromatography type GC
Chromatography system Agilent 6890N
Column Restek corporation Rtx-5Sil MS
MS Type EI
MS instrument type GC-TOF
MS instrument name Leco Pegasus III GC TOF
Ion Mode POSITIVE
Units Peak Area

Chromatography:

Chromatography ID:CH000053
Methods Filename:SOP_GCTOF11082012.pdf
Instrument Name:Agilent 6890N
Column Name:Restek corporation Rtx-5Sil MS
Column Temperature:50-330oC
Flow Rate:1ml/min
Injection Temperature:50C ramped to 250C by 12C s-1
Sample Injection:0.5uL
Oven Temperature:50C for 1 min, then ramped at 20C min-1 to 330C, held constant for 5 min
Transferline Temperature:230C
Washing Buffer:Ethyl Acetate
Sample Loop Size:30 m length x 0.25 mm internal diameter
Randomization Order:Excel generated
Chromatography Type:GC

MS:

MS ID:MS000065
Analysis ID:AN000084
Instrument Name:Leco Pegasus III GC TOF
Instrument Type:GC-TOF
MS Type:EI
Ion Mode:POSITIVE
Ion Source Temperature:250°C
Ionization Energy:-70eV
Mass Accuracy:Nominal
Source Temperature:250°C
Scanning Range:80-500 Da
Acquisition Parameters File:SOP_GCTOF11082012.pdf
Processing Parameters File:Data_Dictionary_Fiehn_Laboratory_GCTOF_MS_primary_metabolism_09-27-2013_general.pdf