Return to study ST001428 main page

MB Sample ID: SA120272

Local Sample ID:C-1ZZY5-S-01_002
Subject ID:SU001502
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Age Or Age Range:ages 2 to 18 years

Select appropriate tab below to view additional metadata details:


Subject:

Subject ID:SU001502
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Age Or Age Range:ages 2 to 18 years

Factors:

Local Sample IDMB Sample IDFactor Level IDLevel ValueFactor Name
C-1ZZY5-S-01_002SA120272FL014873SerumSample type

Collection:

Collection ID:CO001497
Collection Summary:Serum obtained from blood vein draw for children, which are acquired near in time to liver biopsy, for the subject in the NASH CRN cohort. Blood samples were collected in red top vacutainer tubers and sit upright for 30 minutes for clot formation, followed by centrifugation and removal of 500uL aliquots of serum to 1.8mL freezer tubes and stores in a -70C freezer. The samples were shipped on dry ice from the clinical centers to CHEAR URR at Emory University.
Sample Type:Blood (serum)
Storage Conditions:Described in summary

Treatment:

Treatment ID:TR001517
Treatment Summary:Samples were received frozen in aliquouts of <250uL. Freeze-thaw history for study samples prior to receipt by the Emory URR is provided in the Study Design section. Prior to analysis, samples were thawed and prepared for HRM analysis using the standard protocols in the Sample Preparation section.

Sample Preparation:

Sampleprep ID:SP001510
Sampleprep Summary:Samples were prepared for metabolomics analysis using established methods(Johnson et al. (2010). Analyst; Go et al. (2015). Tox Sci). Prior to analysis, plasma aliquots were removed from storage at -80 degrees C and thawed on ice. Each cryotube was then vortexed briefly to ensure homogeneity, and 50 microliters was transferred to a clean microfuge tube. Immediately after, the plasma was treated with 100 microliters of ice-cold LC-MS grade acetonitrile (Sigma Aldrich) containing 2.5 microliters of internal standard solution with eight stable isotopic chemicals selected to cover a range of chemical properties. Following addition of acetonitrile, urine was equilibrated for 30 min on ice, upon which precipitated proteins were removed by centrifuge (14,000 rpm at 4 degrees C for 10 min). The resulting supernatant (100 microliters) was removed, added to a low volume autosampler vial and maintained at 4 degrees C until analysis (<22 h).
Sampleprep Protocol ID:HRM_SP_082016_01
Sampleprep Protocol Filename:EmoryUniversity_HRM_SP_082016_01.pdf
Sampleprep Protocol Comments:Date effective: 30 July 2016
Extraction Method:2:1 acetonitrile: sample followed by vortexing and centrifugation

Combined analysis:

Analysis ID AN002387 AN002388
Analysis type MS MS
Chromatography type HILIC Reversed phase
Chromatography system Dionex UltiMate 3000 Dionex UltiMate 3000
Column Waters XBridge BEH Amide XP HILIC (50 x 2.1mm,2.5um) with Thermo Accucore HILIC guard Higgins endcapped C18 stainless steel (50 x 2.1mm,3um) with Thermo Accucore HILIC guard
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE NEGATIVE
Units peak area peak area

Chromatography:

Chromatography ID:CH001754
Chromatography Summary:The HILIC column is operated parallel to reverse phase column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10-port and 6-port switching valves. During operation of HILIC separation method, the MS is operated in positive ion mode and 10 microliters of sample is injected onto the HILIC column while the reverse phase column is flushing with wash solution. Flow rate is maintained at 0.35 mL/min until 1.5 min, increased to 0.4 mL/min at 4 min and held for 1 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 2% formic acid (v/v) in LC-MS grade water. Initial mobile phase conditions "are 22.5% A, 75% B, 2.5% C hold for 1.5 min, with linear gradient to 77.5% A," "20% B, 2.5% C at 4 min, hold for 1 min, resulting in a total analytical run time" "of 5 min. During the flushing phase (reverse phase analytical separation), the HILIC column is equilibrated with a wash solution of 77.5% A, 20% B, 2.5% C.
Methods ID:2% formic acid in LC-MS grade water
Methods Filename:20200407_posHILIC120kres5min_ESI_c18negwash.meth
Chromatography Comments:Triplicate injections for each chromatography mode. Waters XBridge BEH Amide XP HILIC ( 2.1mm x 50mm x 2.5um) with Thermo Accucore HILIC guard column
Instrument Name:Dionex UltiMate 3000
Column Name:Waters XBridge BEH Amide XP HILIC (50 x 2.1mm,2.5um) with Thermo Accucore HILIC guard
Column Temperature:60C
Sample Injection:10 uL
Solvent A:100% water
Solvent B:100% acetonitrile
Analytical Time:5 min
Sample Loop Size:15 uL
Sample Syringe Size:100 uL
Chromatography Type:HILIC
  
Chromatography ID:CH001755
Chromatography Summary:The C18 column is operated parallel to the HILIC column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10-port and 6- port switching valves. During operation of the C18 method, the MS is operated in negative ion mode and 10 uL of sample is injected onto the C18 column while the HILIC column is flushing with wash solution. Flow rate is maintained at 0.4 mL/min until 1.5 min, increased to 0.5 mL/min at 2 min and held for 3 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 10mM ammonium acetate in LC-MS grade water. Initial mobile phase conditions are 60% A, 35% B, 5% C hold for 0.5 min, with linear gradient to 0% A, 95% B, 5% C at 1.5 min, hold for 3.5 min, resulting in a total analytical run time of 5 min. During the flushing phase (HILIC analytical separation), the C18 column is equilibrated with a wash solution of 0% A, 95% B, 5% C until 2.5 min, followed by an equilibration solution of 60% A, 35% B, 5% C for 2.5 min.
Methods ID:10mM ammonium acetate in LC-MS grade water
Methods Filename:20200407_negC18120kres5min_ESI_HILICposwash.meth
Chromatography Comments:Higgins endcapped C18 stainless steel column. (2.1mm x 50mm x 3μm) with Thermo Accucore HILIC guard column. Product #TS-0521-C183; Thermo Accucore C18 guard column with holder, Product #17126-014005
Instrument Name:Dionex UltiMate 3000
Column Name:Higgins endcapped C18 stainless steel (50 x 2.1mm,3um) with Thermo Accucore HILIC guard
Column Temperature:60C
Flow Rate:0.4 mL/min for 1.5 min; linear increase to 0.5 mL/min at 2 min held for 3 min
Sample Injection:10 uL
Solvent A:100% water
Solvent B:100% acetonitrile
Analytical Time:5 min
Sample Loop Size:15 uL
Sample Syringe Size:100 uL
Chromatography Type:Reversed phase

MS:

MS ID:MS002229
Analysis ID:AN002387
Instrument Name:Thermo Q Exactive HF hybrid Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:none
Ion Mode:POSITIVE
Capillary Temperature:250C
Collision Gas:N2
Dry Gas Flow:45
Dry Gas Temp:150C
Mass Accuracy:< 3ppm
Spray Voltage:3500
Activation Parameter:5.00E+05
Activation Time:118ms
Interface Voltage:S-Lens RF level= 55
Analysis Protocol File:EmoryUniversity_HRM_QEHF-MS_092017_v1.pdf
  
MS ID:MS002230
Analysis ID:AN002388
Instrument Name:Thermo Q Exactive HF hybrid Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:none
Ion Mode:NEGATIVE
Capillary Temperature:250C
Collision Gas:N2
Dry Gas Flow:45
Dry Gas Temp:150C
Mass Accuracy:< 3ppm
Spray Voltage:-4000
Activation Parameter:5.00E+05
Activation Time:118ms
Interface Voltage:S-Lens RF level= 55
Analysis Protocol File:EmoryUniversity_HRM_QEHF-MS_092017_v1.pdf