MB Sample ID: SA155071

Local Sample ID:s24_1
Subject ID:SU001759
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Age Or Age Range:19-87
Gender:Male and female
Human Race:Caucasian

Select appropriate tab below to view additional metadata details:


Subject:

Subject ID:SU001759
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Age Or Age Range:19-87
Gender:Male and female
Human Race:Caucasian

Factors:

Local Sample IDMB Sample IDFactor Level IDLevel ValueFactor Name
s24_1SA155071FL018269TGLabel
s24_1SA155071FL018269b3Batch

Collection:

Collection ID:CO001752
Collection Summary:All morning fasting urine samples were collected in sterilized plastic sealed falcons for this study and were frozen in the refrigerator at -80 °C. All the samples were defrosted at room temperature prior the analysis. The quality control (QC) sample was prepared from a mix of all urine samples that were participated in this research in equal volume, then were filled into separate containers and were stored together with other samples. A new aliquot (from individual container) of QC was used in each new day of analysis.
Sample Type:Urine
Storage Conditions:-80℃

Treatment:

Treatment ID:TR001772
Treatment Summary:Creatinine level, age, sex and clinical diagnosis were collected for study samples as meta-data. The study included urine samples from patients with bladder cancer (TG) and healthy controls (CG) matched with the TG group for age and sex.

Sample Preparation:

Sampleprep ID:SP001765
Sampleprep Summary:Briefly, dilute and shoot technique was utilized. Prior to analysis, samples were defrosted at room temperature and vortexed vigorously and centrifuged at 16000 rpm for 15 min, then aliquots of the supernatant were diluted in 5-fold by deionized water with a final papaverine concentration of 0.2 µg mL-1 (IS), vortexed vigorously and centrifuged at 16000 rpm for 15 min and then aliquots of supernatant were placed into vials.
Processing Storage Conditions:Room temperature
Extract Storage:Room temperature

Combined analysis:

Analysis ID AN002746
Analysis type MS
Chromatography type Reversed phase
Chromatography system Shimadzu 20AD
Column Waters Acquity BEH C18 (100 x 2mm,1.7um)
MS Type ESI
MS instrument type Other
MS instrument name Shimadzu LCMS-IT-TOF
Ion Mode POSITIVE
Units area

Chromatography:

Chromatography ID:CH002028
Chromatography Summary:The apparatus consisted of a two LC-20AD pumps (Shimadzu, Japan), an automatic sample injector, coupled on-line with an LCMS-IT-TOF (Shimadzu, Japan) mass spectrometer with an electrospray ionization source. The HPLC separation was conducted on a 2.1 × 100 mm Waters Acquity UPLC BEH; 1.7 μm column (Waters, Milford MA, USA) with 2.1 × 5 mm UHPLC Guard 3PK Zorbax SB-C18 1.8 μm guard column (Agilent Technologies, USA). Two solvents were used: (A) deionized water with 0,1% formic acid, and (B) MeCN. After a 5 min hold at 5% solvent B, elution was performed according to the following conditions: from 5% B to 40% B in 20 min; to 95% B in 7 min; keeping constant for 10 min, to 5% B in 3 min, followed by 5 min of equilibration on the initial conditions. Total analysis time was 50 min. Column temperature was 35˚C. Autosampler tray temperature was 4˚C.
Instrument Name:Shimadzu 20AD
Column Name:Waters Acquity BEH C18 (100 x 2mm,1.7um)
Column Pressure:300-500 bar
Column Temperature:35
Flow Gradient:non-linear
Flow Rate:0.3 ml/min
Injection Temperature:5
Internal Standard:papaverine
Sample Injection:5 mkl
Solvent A:100% water; 0.1% formic acid
Solvent B:100% acetonitrile
Analytical Time:50 min
Chromatography Type:Reversed phase

MS:

MS ID:MS002543
Analysis ID:AN002746
Instrument Name:Shimadzu LCMS-IT-TOF
Instrument Type:Other
MS Type:ESI
MS Comments:The apparatus consisted of a two LC-20AD pumps (Shimadzu, Japan), an automatic sample injector, coupled on-line with an LCMS-IT-TOF (Shimadzu, Japan) mass spectrometer. The column effluent was analyzed by ESI-MS. MS detection was operated in positive ionisation mode with the following parameters: detector voltage, 1.53 kV; interface voltage, +4.5 kV; CDL (curve desolvation line) temperature, 200 ˚C; block heater temperature, 200 ˚C; nebulizing gas flow (He), 1.5 L/min; drying gas pressure, 100 kPa. Full scan MS data were acquired in the range of 80–850 m/z (ion accumulation time, 30 ms; event time, 300 ms; repeat = 3.
Ion Mode:POSITIVE
Capillary Voltage:+4.5 kV
Dry Gas Flow:100 kPa
Helium Flow:1.5 ml/min
Ion Source Temperature:200 ˚C
Ionization:positive
Mass Accuracy:15 ppm
Source Temperature:200 ˚C
Cdl Temperature:200 ˚C